Press
START
.
Upon completing the measurement read the result. The test can be finished
before the defined test time duration by pressing
STOP
.
Positive test result: R
PE
< LIMIT
Negative test result: R
PE
> LIMIT
Notes:
- Test circuit is electrically isolated from the mains and from mains' PE lead.
- Continous test is available only for 200mA test current.
4.3 Measurement of insulation resistance
Note:
- For Class I appliances, the previous R
PE
test has to have positive result.
Press
Riso
.