4.5 Measurement of substitute leakage current
Notes:
- For Class I appliances, the previous R
PE
test has to have positive result.
- I
SUB
current is measured at <50 V voltage and its value is rescaled to the nominal mains voltage
value set in the menu (see 3.3.5). The voltage is applied between shorted L ad N, and PE. The resi-
stance of the measuring circuit is 2k
Ω
.
Press
Isub
.
Before starting the measurement, similarily to the measurement of PE continuity test (see 4.2 of
this manual), following parameters must be set:
test duration, limit
.
For Class I, connect the mains plug of the tested appliance into the test socket of the tester.
For Class II and accessible parts disconnected from PE in Class I, connect the probe to
T2
terminal socket and touch the accessible parts of the tested device.