somnotec
GmbH D Hennigsdorf
QUISI
®
Version 2.0 (Stand: 28.06.2001 ) tl 02.02. / rev. 0103
24.02.2003
Seite 6 von 45
1.3.3. Disposal notes
•
The recorder is to be disposed with attention as
special trash,
as it contains a Nickel metal
hydride battery. Disposal as normal domestic trash is not permitted.
•
The single-use electrodes can be disposed via the normal domestic refuse / trash.
1.4. General specifications
•
Real-time processing of EEG by the recorder (feature extraction)
•
On-line recording of the EEG (with the PC connected by fiber optic cable)
•
Frontal single-channel evaluation of the EEG-signal
•
Adaptive classification by neural networks (PC software)
•
Signal input range 250µV
•
Sampling frequency 128Hz
•
Off-line processing of the sleep EEG in epochs of 30 seconds
•
24 hours stand-by
•
Alarm function
•
Fully-automated battery-recharging takes between half an hour and 4 hours
•
Power supply: 4 x 1.2 V of NiMH batteries (1300 mAh)
•
Accessories power pack: input 230V~, output 9V=/667mA
•
Loading- and application temperature: 5°C to +40°C
•
Storage temperature: -20°C to +40°C
The QUISI
®
software runs under Windows 95, 98, NT (3.51 or higher), 2000 or ME.