Remote Radio Link User Guide
Page 28
4.3 Barometric Compensation
Submersed Leveloggers measure total or absolute pressure (water column equi barometric pressure).
In order to accurately determine the true changes in water level only, Levelogger data must be barometrically
compensated. Compensation simply involves subtraction of the barometric reading from the corresponding
Levelogger reading.
Levelogger data can be barometrically compensated using data from a local Barologger (one Barologger can
cover all Leveloggers within a 20 mile/30 km radius or per 1000 ft/300 m change in elevation) using the
Barometric Compensation application in the STS /RRL Administrator.
The Levelogger and Barologger must have the same time stamp in order for the compensation to be
successful.
A Barologger Edge or Barologger Gold can be used to compensate the data from any version of
Levelogger.
Note:
Alternatively, RRL data files can be exported using the STS/RRL Administrator as *.xle or *.lev files and opened in
Levelogger Software for barometric compensation. See the Levelogger User Guide for more details.
To begin the compensation, click the
Barometric Compensation
tab on the STS/RRL Administrator window.
In the Barometric Compensation Window, from the “Enable Barometric Compensation” list, check the RRL
Station that you wish to compensate, and RRL Station (or STS Site) that will provide the Barologger for the
compensation.
In the “Select Barologger from Site” section, select the RRL Station (or STS Site) that contains the Barologger
you wish to use for the compensation, and identify the Barologger by its position (Datalogger ID) on the RRL
Station (or STS Site).
The Levelogger and Barologger must have the same time stamp in order for the compensation to be
successful.
To account for different deployment altitudes in your compensation, you can choose to enter an altitude for
your Barologger and each datalogger.
Click OK to complete the barometric compensation set up.
To view compensated data, see Section 7.2.1
Figure 4-23 Barometric Compensation Window