Remote Radio Link User Guide
Page 20
Figure 4-11 LTC Levelogger Measurement Parameters
The Conductivity Channel includes the following parameters:
• Identification
describes the measurement parameter of the channel and has already been
configured as ‘Conductivity’ .The Identification field will be the channel heading, data column
heading and graph line name when viewing the data.
• Unit
refers to the channel’s units of measurement. There are two units of measure available for the
user to select: mS/cm or μS/cm.
• The Temperature Coefficient
(LTC Levelogger Junior only) field allows you to choose the actual
conductivity (‘Conductivity’) at the current temperature or select Specific Conductance (‘Spec.
Cond’.)measurement in which the conductivity reading is temperature compensated to the standard
of 25ºC. The Temperature Coefficient default is 2.00 for Specific Conductance readings. The
Temperature Coefficient should not be adjusted, unless you know the value specific to the solution
you are measuring.