Levelogger Series User Guide
Page 42
When a single-point calibration is successful, the process is complete. Once the calibration is complete for
the first solution of a multipoint calibration, a message will appear stating that the calibration was successful.
When you select 'Okay', the Calibration Wizard will go back to the beginning of Step 3 to start the rinsing
process for the next solution selected. The calibration process will proceed automatically until completed for
all solutions.
If a failure occurs at any point during calibration, a message will appear asking you to clean and check
your probe, then start the calibration process over for that current solution by selecting 'Yes' (Figure 6-7).
If performing a multipoint calibration, any previous successful calibration points will remain.
Figure 6-7 Levelogger Conductivity Calibration Warning 1
If a second failure occurs during calibration, a warning message will appear indicating that your probe may still
be dirty, or damaged. This may occur if your Solinst LTC Levelogger conductivity sensor has been affected
by dirt, mineral build-up, etc., so it no longer responds like it did when it was first factory calibrated. This step
allows your sensor to be calibrated within a wider range of the standard solution value.
Figure 6-8 Levelogger Conductivity Calibration Warning 2