background image

Model 464 Electronic Pump Control Unit User Guide

Page 14

5.0  Maintenance

Moisture in the Model 464 Electronic Pump Control Unit can shorten the 

lifetime of some of its components. After each sampling session, compressed air 

or nitrogen should be sent through the Pump Control Unit to ensure that any 

build-up of moisture is purged and pushed out of the Pump Control Unit, as well 

as through the Supply and Drive Lines.

Connect the compressed air/nitrogen gas source and let it cycle through the 

Pump Control Unit with the drive line disconnected at the pneumatic pump. Set 

a 10 second "Drive Time" and 2 second "Vent Time". This will create a pulse of 

air to purge the system components. Maintain this purging cycle for 10 minutes.

If you are using the 464 External Filter, use the draining valve as necessary to 

prevent water from entering the 464 Electronic Pump Control Unit. 

Of note, sources of moisture may include:

1. The air compressor − follow the manufacture's suggestions for routine 

maintenance for draining the air tank from a build-up of moisture.

2. Sample backflow − best practice is to maintain the Control Unit at a higher 

elevation than the Drive Line connection at the wellhead. This height 

difference reduces the potential for backflow into the Control Unit.

3. Precipitation  − avoid direct exposure to precipitation or water splashing;  

 

 

− ensure that your Control Unit is stored dry, inside and out.

4.  Pump damage − leaks in tubing, a ruptured bladder, or damaged o-rings 

can cause backflow. Regularly monitor pump condition and replace parts as 

required. 

 

  NOTE

If moisture is observed exiting the 
diffuser on the Drive Line exhaust 
fitting when the pump is venting, 
remove the pump from the well 
and check for damage (o-rings, 
bladder, lines).  

Содержание 464 Mk3

Страница 1: ...Model 464 Mk3 Electronic Pump Control Unit User Guide September 24 2019 109652 Rev C...

Страница 2: ......

Страница 3: ...Rates 5 2 5 1 Saved Settings 5 Edit Settings 6 Delete Settings 6 2 5 2 Create Setting 7 2 6 Automatic Drive Vent Cycles 7 2 7 Manual Drive Vent Cycles 8 2 8 Battery Replacement 8 3 0 Pumping Instruct...

Страница 4: ......

Страница 5: ...4 9 23 x 26 x 12 cm Drive Vent Time Range 1 999 seconds Maximum Site Name Length 16 characters upper or lowercase and numeric Memory Capacity 99 user flow rates in non volatile FRAM 3 preset in Flash...

Страница 6: ...ghlighted menu item also toggles between lower and uppercase letters and pressure units Press and hold for at least 3 seconds to turn the Control Unit off However this will not work in the Contrast Me...

Страница 7: ...1 905 873 2255 2 1 1 Main Menu There are four main menu items for the Electronic Control Unit Figure 2 2 Preset Flow Rates allows you to select a flow rate with predefined drive and vent cycles Low F...

Страница 8: ...V2 000 Solinst Canada Ltd 1 905 873 2255 www solinst com About ECU 464 125 4 59 Min 99 Set LCD Contrast 50 4 59 Min 99 2 4 Preset Flow Rates Preset Flow Rates have predefined drive and vent times they...

Страница 9: ...tings Selecting Saved Settings will display a list of all previously created user defined flow rates Figure 2 9 User Defined Saved Settings Use the cursor keys to navigate through the settings To rear...

Страница 10: ...on the Unit Pressing OK toggles between psi and KPa The pressure is automatically converted to the new units The displayed pressure does not control the actual pressure it is only a reminder of the pr...

Страница 11: ...99 2 6 Automatic Drive Vent Cycles Select the desired setting To start the drive vent cycles use the cursor key to highlight Start and press OK When the Control Unit is running the LCD will display th...

Страница 12: ...n Battery Enclosure Pressure Gauge If the battery level is low the battery warning will be displayed after attempting to start a drive vent cycle Figure 2 16 The battery icon will also be flashing Pre...

Страница 13: ...reel The Drive Line supplied with the Quick Exhaust fitting should always be used 4 Press any button on the keypad to turn the Control Unit on 5 Select the desired pumping settings 6 Set the compress...

Страница 14: ...utton on the Electronic Control Unit Count the seconds until the bubbles stop forming in the water This count is your Drive Time as when the air bubbles stop the bladder is fully compressed Bladder Pu...

Страница 15: ...Drive Time is 40 of this count i e the full count is the amount of time is takes to evacuate the entire pump system of water using drive gas 40 of that count means the drive gas will not enter the pum...

Страница 16: ...increase the vent time then re adjust the drive time to obtain the highest flow rate Double Valve Pumps Select and start your desired flow rate from the User Flow Rates menu If a higher flow rate is r...

Страница 17: ...to use the External Filter The External Filter connects to the 464 Supply Line and reduces the risk of moisture entering the Control Unit See separate operating and installation instructions for the 1...

Страница 18: ...r use the draining valve as necessary to prevent water from entering the 464 Electronic Pump Control Unit Of note sources of moisture may include 1 The air compressor follow the manufacture s suggesti...

Страница 19: ...ged with compressed air after use as described in Section 5 0 then stored in a dry condition Ensure the 464 External Filter is dry before storage If the Pump Control Unit is going to be stored for lon...

Страница 20: ...Ltd 35 Todd Road Georgetown ON L7G 4R8 Fax 1 905 873 1992 800 516 9081 Tel 1 905 873 2255 800 661 2023 instruments solinst com www solinst com High Quality Groundwater and Surface Water Monitoring In...

Отзывы: