Model 464 Electronic Pump Control Unit User Guide

Page 9

3.0  Pumping Instructions

3.1  Preparation

•  The Control Unit is shipped without the batteries in the housing.   

  Install these before operating the unit. The Control Unit uses four  

  (4) AA alkaline batteries.
•  Do not let water get inside the Control Unit. Always position the  

  Control Unit physically higher than the sample discharge and  

  wellhead. This helps avoid the siphoning effect, where gravity back- 

  flow of sample water can enter the Control Unit and cause damage.
•  Do not release any pressure from the compressed gas supply until all  

  preparatory steps are complete.

Figure 3-1 Pumping Set Up

464 

Electronic Control Unit 

(125 psi)

Drive

 

Line

Sample

 

Line

Supply 

Line

12 Volt 

Compressor

Sample 

Bottle

3.2  Pumping Set Up 

1.  Connect the Supply Line to the Air In fitting on the Control Unit. Attach the 

other end to the compressed gas supply source.

2.  Ensure your pump is installed to the desired pumping depth with the sample 

and drive tubing properly connected.

3.  Connect the Drive Line from the Air Out fitting to the quick connect on 

the wellhead manifold (or air supply connection on a reel). The Drive Line 

supplied, with the Quick Exhaust fitting, should always be used.

4.  Press any button on the keypad to turn the Control Unit on.
5.  Select the desired pumping settings.
6.  Set the compressed gas supply regulator to no more than 150 psi for the 

125 psi Control Unit, or 300 psi for the 250 psi Control Unit.

7.  Start the selected drive/vent cycle.
8.  Allow several cycles for water to reach the surface, then adjust the settings 

as necessary.

Quick 

Exhaust

 

  NOTE

Always follow local health and 
safety practices.  Work safely!

 

  NOTE

Keep Control Unit physically higher 
than the sample discharge and 
wellhead.

 

  NOTE

While the drive/vent cycles are in 
progress it is not possible to edit the 
displayed site information or exit 
the site menu when using a Saved 
Setting.  

 

  NOTE

The Pressure Regulator on the 
Control Unit may hiss during 
operation; this is normal.   

 

  NOTE

You should always use the Drive 
Line supplied with the Control 
Unit, as it uses the Quick Exhaust 
fitting, which is required for proper 
operation.

 

  NOTE

When using a Solinst 12 Volt 
Compressor, or any other air 
compressor, it is recommended 
to use the 464 External Filter. See 
Section 4 for more details. 

 

  NOTE

The Control Unit will turn off after 
five minutes of being idle, provided 
the Unit is not cycling.  To turn the 
Unit off, hold down the OK button 
for at least 3 seconds.  

Содержание 464 Mk3

Страница 1: ...Model 464 Mk3 Electronic Pump Control Unit User Guide September 24 2019 109652 Rev C...

Страница 2: ......

Страница 3: ...Rates 5 2 5 1 Saved Settings 5 Edit Settings 6 Delete Settings 6 2 5 2 Create Setting 7 2 6 Automatic Drive Vent Cycles 7 2 7 Manual Drive Vent Cycles 8 2 8 Battery Replacement 8 3 0 Pumping Instruct...

Страница 4: ......

Страница 5: ...4 9 23 x 26 x 12 cm Drive Vent Time Range 1 999 seconds Maximum Site Name Length 16 characters upper or lowercase and numeric Memory Capacity 99 user flow rates in non volatile FRAM 3 preset in Flash...

Страница 6: ...ghlighted menu item also toggles between lower and uppercase letters and pressure units Press and hold for at least 3 seconds to turn the Control Unit off However this will not work in the Contrast Me...

Страница 7: ...1 905 873 2255 2 1 1 Main Menu There are four main menu items for the Electronic Control Unit Figure 2 2 Preset Flow Rates allows you to select a flow rate with predefined drive and vent cycles Low F...

Страница 8: ...V2 000 Solinst Canada Ltd 1 905 873 2255 www solinst com About ECU 464 125 4 59 Min 99 Set LCD Contrast 50 4 59 Min 99 2 4 Preset Flow Rates Preset Flow Rates have predefined drive and vent times they...

Страница 9: ...tings Selecting Saved Settings will display a list of all previously created user defined flow rates Figure 2 9 User Defined Saved Settings Use the cursor keys to navigate through the settings To rear...

Страница 10: ...on the Unit Pressing OK toggles between psi and KPa The pressure is automatically converted to the new units The displayed pressure does not control the actual pressure it is only a reminder of the pr...

Страница 11: ...99 2 6 Automatic Drive Vent Cycles Select the desired setting To start the drive vent cycles use the cursor key to highlight Start and press OK When the Control Unit is running the LCD will display th...

Страница 12: ...n Battery Enclosure Pressure Gauge If the battery level is low the battery warning will be displayed after attempting to start a drive vent cycle Figure 2 16 The battery icon will also be flashing Pre...

Страница 13: ...reel The Drive Line supplied with the Quick Exhaust fitting should always be used 4 Press any button on the keypad to turn the Control Unit on 5 Select the desired pumping settings 6 Set the compress...

Страница 14: ...utton on the Electronic Control Unit Count the seconds until the bubbles stop forming in the water This count is your Drive Time as when the air bubbles stop the bladder is fully compressed Bladder Pu...

Страница 15: ...Drive Time is 40 of this count i e the full count is the amount of time is takes to evacuate the entire pump system of water using drive gas 40 of that count means the drive gas will not enter the pum...

Страница 16: ...increase the vent time then re adjust the drive time to obtain the highest flow rate Double Valve Pumps Select and start your desired flow rate from the User Flow Rates menu If a higher flow rate is r...

Страница 17: ...to use the External Filter The External Filter connects to the 464 Supply Line and reduces the risk of moisture entering the Control Unit See separate operating and installation instructions for the 1...

Страница 18: ...r use the draining valve as necessary to prevent water from entering the 464 Electronic Pump Control Unit Of note sources of moisture may include 1 The air compressor follow the manufacture s suggesti...

Страница 19: ...ged with compressed air after use as described in Section 5 0 then stored in a dry condition Ensure the 464 External Filter is dry before storage If the Pump Control Unit is going to be stored for lon...

Страница 20: ...Ltd 35 Todd Road Georgetown ON L7G 4R8 Fax 1 905 873 1992 800 516 9081 Tel 1 905 873 2255 800 661 2023 instruments solinst com www solinst com High Quality Groundwater and Surface Water Monitoring In...

Отзывы: