2.0  Control Unit Operation

2.1  Start Up

Press any button on the keypad to turn the Control Unit on.  When first starting 

the Control Unit with new batteries, or restarting after replacing the batteries, 

the Control Unit will perform a self-test to identify any faults that may exist.  

The start-up screen is shown first (Figure 2-1 Start-up Screen After Replacing 

Batteries); following a short pause the main menu will be displayed.  

Figure 2-1 Start-up Screen After Replacing Batteries

ECU 464 V2.000

Copyright (c) 2016

Solinst Canada Ltd.

+1 (905) 873-2255

2.1.1  Main Menu

There are four main menu items for the Electronic Control Unit (Figure 2-2).  

Preset Flow Rates: allows you to select a flow rate with predefined drive and vent 

cycles (Low Flow, Medium Flow, or High Flow).

User Flow Rates: allows you to create user-defined flow rates, save flow rates, 

select and edit saved flow rates. 

LCD Contrast: enables you to adjust the brightness of the display.

About: will display information about the Control Unit, including firmware 

version and Solinst contact information. 

Figure 2-2 Model 464 Main Menu

User Flow Rates
LCD Contrast
About

ECU 464-125 Main

Preset Flow Rates

4:59 Min

99%

  NOTE

The Control Unit does not come 
with the batteries installed.

 

  NOTE

Turning the Control Unit on, 
displays the last screen shown at 
the time of shut-down.

Содержание 464 Mk3

Страница 1: ...Model 464 Mk3 Electronic Pump Control Unit User Guide September 24 2019 109652 Rev C...

Страница 2: ......

Страница 3: ...Rates 5 2 5 1 Saved Settings 5 Edit Settings 6 Delete Settings 6 2 5 2 Create Setting 7 2 6 Automatic Drive Vent Cycles 7 2 7 Manual Drive Vent Cycles 8 2 8 Battery Replacement 8 3 0 Pumping Instruct...

Страница 4: ......

Страница 5: ...4 9 23 x 26 x 12 cm Drive Vent Time Range 1 999 seconds Maximum Site Name Length 16 characters upper or lowercase and numeric Memory Capacity 99 user flow rates in non volatile FRAM 3 preset in Flash...

Страница 6: ...ghlighted menu item also toggles between lower and uppercase letters and pressure units Press and hold for at least 3 seconds to turn the Control Unit off However this will not work in the Contrast Me...

Страница 7: ...1 905 873 2255 2 1 1 Main Menu There are four main menu items for the Electronic Control Unit Figure 2 2 Preset Flow Rates allows you to select a flow rate with predefined drive and vent cycles Low F...

Страница 8: ...V2 000 Solinst Canada Ltd 1 905 873 2255 www solinst com About ECU 464 125 4 59 Min 99 Set LCD Contrast 50 4 59 Min 99 2 4 Preset Flow Rates Preset Flow Rates have predefined drive and vent times they...

Страница 9: ...tings Selecting Saved Settings will display a list of all previously created user defined flow rates Figure 2 9 User Defined Saved Settings Use the cursor keys to navigate through the settings To rear...

Страница 10: ...on the Unit Pressing OK toggles between psi and KPa The pressure is automatically converted to the new units The displayed pressure does not control the actual pressure it is only a reminder of the pr...

Страница 11: ...99 2 6 Automatic Drive Vent Cycles Select the desired setting To start the drive vent cycles use the cursor key to highlight Start and press OK When the Control Unit is running the LCD will display th...

Страница 12: ...n Battery Enclosure Pressure Gauge If the battery level is low the battery warning will be displayed after attempting to start a drive vent cycle Figure 2 16 The battery icon will also be flashing Pre...

Страница 13: ...reel The Drive Line supplied with the Quick Exhaust fitting should always be used 4 Press any button on the keypad to turn the Control Unit on 5 Select the desired pumping settings 6 Set the compress...

Страница 14: ...utton on the Electronic Control Unit Count the seconds until the bubbles stop forming in the water This count is your Drive Time as when the air bubbles stop the bladder is fully compressed Bladder Pu...

Страница 15: ...Drive Time is 40 of this count i e the full count is the amount of time is takes to evacuate the entire pump system of water using drive gas 40 of that count means the drive gas will not enter the pum...

Страница 16: ...increase the vent time then re adjust the drive time to obtain the highest flow rate Double Valve Pumps Select and start your desired flow rate from the User Flow Rates menu If a higher flow rate is r...

Страница 17: ...to use the External Filter The External Filter connects to the 464 Supply Line and reduces the risk of moisture entering the Control Unit See separate operating and installation instructions for the 1...

Страница 18: ...r use the draining valve as necessary to prevent water from entering the 464 Electronic Pump Control Unit Of note sources of moisture may include 1 The air compressor follow the manufacture s suggesti...

Страница 19: ...ged with compressed air after use as described in Section 5 0 then stored in a dry condition Ensure the 464 External Filter is dry before storage If the Pump Control Unit is going to be stored for lon...

Страница 20: ...Ltd 35 Todd Road Georgetown ON L7G 4R8 Fax 1 905 873 1992 800 516 9081 Tel 1 905 873 2255 800 661 2023 instruments solinst com www solinst com High Quality Groundwater and Surface Water Monitoring In...

Отзывы: