-RF: Run/Fill Switch Option for 3G (3 lpm Gear Pump)
Thermocubes with -3G pumps with the RF option come with
a
Run/Fill switch located above the signals connector. This switch
allows the user to run the pump at low speed (Fill) during initial
filling and then, once all air-bubbles are out of the plumbing lines,
increase the pump speed (Run) for normal operation. Failure to run
the pump in the fill mode after initial hook-up or with drained
plumbing lines can result in water being blown out of the tank from
compressed air bubbles in the lines.
Note: The coolant flow rate in the fill mode is approximately 1.3
liters/minute.
-SS: Stainless Steel Plumbing and Tank Option
With this option the standard tank and plumbing are replaced with all
stainless steel tank and plumbing. Note: This configuration option
is only available with gear pumps.
-ST: Stainless Steel Tank Option
With this option the standard tank is replaced with a stainless steel
tank.
-VS: Variable Speed Fan Option
The ThermoCube can be configured with a variable speed fan option
that slows the fan speed down based upon the internal air heat sink
temperature. This option substantially reduces the ThermoCube
noise when the ThermoCube is not running at full cooling. When
full cooling is required, the fans speed increases to full power. This
option can lead to reduced cooling capacity, depending upon the
ambient air temperature and coolant set point. Contact Solid State
Cooling Systems for further details.
-VD: Vibration Dampening Option
This option adds vibration dampening material to interior sheet metal
surfaces for applications such as scanning electron microscopes that
are highly sensitive to vibrations from pumps and fans. This option
is usually combined with the –EF external filer option which
dampens vibrations in the coolant flow.
- RoHS Compliance
At this time we only offer fully RoHS compliant units as a CES
(Custom Engineered Special). Please contact SSCS for more
information.