35952350 (CH01/AA)
1-41
CH CV
(CHannel Calibrate Voltage)
CH
n
CV
x
y
Where:
n
is an integer that defines the channel number, in the range 1
≤
n
≤
4
x
= the measured voltage value, corresponding to an non-calibrated 0V output
y
= the measured voltage value, corresponding to an non-calibrated 10V output
Values
x
and
y
are both in volts, in floating-point format.
Note:
This command applies only to an IMP type 1D with a ‘3595 3Y’ connector block.
Function:
Allows voltage calibration of a specific output channel. This should normally be left to
Solartron Mobrey, or those users with specialist reference equipment.
Response:
Stream 3, 4 characters
C
n
V
e
, where:
n
is the channel number and
e
represents the
calibration response, as shown in Table 1.18.
Equipment:
The calibration equipment required for IMP type 1D is:
a
. An ‘analogue output’ connector block (P/N: 35953Y). This allows the IMP to recognise
the calibrate commands
b
. A stable temperature environment of 20
°
C
±
3
°
C. The IMP should be powered-up and
allowed to stabilise at this temperature, ideally for 24 hours
c
. An accurate multimeter to measure the output channels.
Procedure:
The procedure for calibrating the output voltage of a type 1D IMP is:
1.
Fit all cables to the ‘3595 3Y’ connector block. PCB terminal connectors are as shown
in Figure 1.4.
a. The S-Net cable is connected to the ‘COMMS IN’ terminals. Strictly ob to +
and - to – polarity, and the S (screen) to S connections throughout the network.
b. The multimeter is connected to the channel to be calibrated.
Note: To avoid signal reflections, a terminator connector (P/N: 35900222) must be
connected across the ‘COMMS OUT’ terminals of the last IMP in the system. (The
terminator connector is supplied with the S-Net host interface package.)
2.
Set the rotary address switches on the ‘3595 3Y’ connector block PCB to a number
between 01 and 50.
3.
Insert the ‘3595 3Y’ connector block into the IMP to be calibrated
4.
Power-up the PC.
5.
Turn the calibration mode off by sending the
KA1
command to the IMP.
6.
Using the
CH VO
command, set the channel output voltage first to 0V and second to
10V, and in each case, measure the voltage actually output by the IMP.
7.
Using the measurements obtained in Step 6, send the channel calibrate voltage
command
CH
n
CV
xy
. In this command,
x
is the actual voltage measured when 0V was
requested and
y
is the actual voltage measured when 10V was requested.
8.
Check the calibration response in stream 3 (see Table 1.18 on page 1-42) to ensure
that the IMP channel has been successfully calibrated.
Содержание 3595 4C
Страница 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Страница 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Страница 6: ...vi 35952350...
Страница 7: ......
Страница 8: ......
Страница 10: ...1 ii 35952350...
Страница 12: ...1 iv 35952350...
Страница 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Страница 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Страница 50: ...2 ii 35952350...
Страница 52: ...2 iv 35952350...
Страница 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Страница 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Страница 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Страница 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Страница 127: ......