![Solartron Mobrey 3595 4C Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/solartron-mobrey/3595-4c/3595-4c_user-manual_1314070071.webp)
3595 4C User Manual Part Two IMP Commands
35952350 (CH01/AA)
1-19
ME
(
ME
asure a channel)
ME
n
Where:
n
is an integer which defines the channel number:
•
in the range 1
≤
n
≤
20 for IMP types 1A, 1C, 1H, 1J and 2A, or
•
in the range 1
≤
n
≤
10 for IMP types 1D or 2B
Note:
This command does not apply to type 1D IMP devices.
Function:
Instructions an IMP to take a single measurement on a specified channel. On IMP types
other than the 1H and 1J, the channels set for event capture, skip, or digital output are not
affected by this command, but returns the error message ‘not measured’ instead. However,
the digital channels (19 and 20) on the 1J and 1H IMP types, when set for digital output,
respond to a
ME
command by returning the present status.
ME
asure starts the counting on a single channel set to ‘event count increment’ or ‘event
count totalise’, provided that counting has not already started. An initial result of zero is then
returned. If counting has already started, the result returned is the number of events the IMP
has recorded in that channel so far.
ME
asure does not reset any counters.
Response:
Stream 1, one 4-byte result.
See also:
RM (for the 1H and 1J IMP types)
RE
(
RE
set)
RE
Function:
Sets all IMP settings to their default values – this
normally
assumed on first power-up:
•
CO
ntinuous ‘off’.
•
Not
armed.
•
Defined scan period set to 0ms.
•
KA
libration ‘on’.
•
DR
ift correct ‘on’.
•
External temperature ‘on’.
•
AM
bient temperature reference ‘on’.
•
All channels to ‘skip’.
•
Time-out 2 seconds.
•
Temperatures returned in
°
C.
•
Integration time set to 20ma (FR0).
•
Voltage and current outputs set to default values (see installation guide).
•
Sets scan to be sent over stream 0.
•
Sets digital sample rate to default.
•
Event counters cleared.
•
User conversions cleared and disabled.
•
Both user thermocouples cleared.
•
Both user post linearisations cleared.
•
All channel alarm conditions cleared and disabled.
Response:
None
Содержание 3595 4C
Страница 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Страница 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Страница 6: ...vi 35952350...
Страница 7: ......
Страница 8: ......
Страница 10: ...1 ii 35952350...
Страница 12: ...1 iv 35952350...
Страница 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Страница 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Страница 50: ...2 ii 35952350...
Страница 52: ...2 iv 35952350...
Страница 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Страница 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Страница 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Страница 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Страница 127: ......