![Solartron Mobrey 3595 4C Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/solartron-mobrey/3595-4c/3595-4c_user-manual_1314070085.webp)
3595 4C User Manual Part Two IMP Commands
35952350 (CH01/AA)
1-33
1.3.4 COMMANDS FOR STRAIN GAUGE MEASUREMENTS
CH GA
(set CHannel GAuge factor)
CH
n
GA
g
Where:
n
is an integer that defines the channel number, in the range 1
≤
n
≤
10
g
is an IEEE 754 floating-point number defining the strain gauge factor
Note:
This command applies only to IMP type 1B with strain gauge(s).
Function:
Loads the IMP database with the gauge factor required for a specified channel. A strain
gauge channel can’t perform measurements until it has been
IN
itialised and the gauge factor
sent to the IMP.
Response:
None.
See also:
IN
,
CH
OF
,
LO
and
SA
CH OF
(set CHannel OFfset and initial voltage)
CH
n
OF
o
p
Where:
n is an integer that defines the channel number, in the range 1
≤
n
≤
10
o is an IEEE 754 floating-point number defining the offset in volts
p is an IEEE 754 floating-point number defining the initial gauge voltage
Note:
This command applies only to IMP type 1B with strain gauge(s).
Function:
Sets the strain gauge offset and initial voltage values used by an IMP to calculate strain
results. These values will have been returned by the
IN
itialise command and should be
stored in the PC. After an IMP has been powered-down, this command can be used to set up
a strain gauge channel to the original condition set by the previous
IN
command.
This must be used in conjunction with the
CH
GA
command before the strain gauge can
return data.
Response:
None.
IN
(INitialise strain gauge parameters)
IN
n
Where:
n
is an integer that defines the channel number, in the range 1
≤
n
≤
10
Note:
This command applies only to IMP type 1B with strain gauge(s).
Function:
Before a strain gauge channel can be used, the IMP must first know the initial voltage of a
gauge connected to a specified channel, to store this, and also transmit this data to the PC.
The IMP must then be informed of the gauge factor using the CH GA command. The IMP
Содержание 3595 4C
Страница 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Страница 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Страница 6: ...vi 35952350...
Страница 7: ......
Страница 8: ......
Страница 10: ...1 ii 35952350...
Страница 12: ...1 iv 35952350...
Страница 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Страница 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Страница 50: ...2 ii 35952350...
Страница 52: ...2 iv 35952350...
Страница 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Страница 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Страница 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Страница 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...