Chapter 7: Respiration
Operation Manual SurgiVet
®
V3404 Plus
7-1
WARNING! This monitor is not for apnea detection. The monitor has not been tested or validated for use
in apnea detection.
Theory of Operation
The V3404 Plus uses a thermistor mounted in the Luer–lock fitting to measure respiration. The thermistor is
inserted into the tracheal tube adapter, to place sensor in the middle of the air stream. The sensor is heated and
detects breath by cooler air moving across the sensor. This enables you to get respiration readings on lower tidal
volume patients.
Respiration Menu
The Respiration menu allows the user to view and/or change monitoring settings.
Figure 7.1: Respiration Menu
Using the Respiration Parameter
If not familiar with this monitor and the Respiration parameter, follow this chapter’s sections in order:
Setting Up Respiration Monitoring
Ensure the respiration parameter is enabled.
RESP MONITOR
Indicates whether the respiration parameter is enabled or disabled. Allows the respiration parameter to be
turned on or off.
The respiration parameter can be turned on or off in the
RESPIRATION
menu. To ensure the respiration
parameter is on, press the
b
key. Use the ARROW (
o n
) keys to highlight the
RESPIRATION
option.
Press the
b
key again to enter the respiration menu. Use the ARROW (
o n
) keys to highlight the
RESP
MONITOR
option, and confirm the option is set to
ON
.
Chapter 7: Respiration
Sp02 SENSOR!
SENSITIVITY SENS 100
RESP MONITOR ON
[EXIT] (respiration menu)
n
Содержание SurgiVet V3404 Plus
Страница 2: ......
Страница 8: ...Table of Contents vi Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 12: ...Warranty and Service Information Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 20: ...Chapter 1 Introduction 1 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 50: ...Chapter 6 Oximetry 6 10 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 64: ...Chapter 10 Serial Output 10 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 66: ...Chapter 11 Analog Output 11 2 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 84: ...Appendix B Revision History B 2 Autocorr Plus Operation Manual This page is intentionally left blank ...
Страница 85: ......