Chapter 6: Oximetry
6-4
Operation Manual SurgiVet
®
V3404 Plus
Adjusting or Viewing the Averaging Settings
SpO
2
averaging is the number of pulse beats over which the SpO
2
value is averaged. Rate averaging is the
number of seconds over which the pulse rate is averaged. Three averaging settings are available: Normal, Fast,
and Slow (4, 8, 16 beats).
To adjust or view the averaging settings, do the following:
Select the
OXIMETER
item from the Main menu.
Select the
AVERAGING
item.
Press the
b
key to select the item. Use the ARROW (
no
) keys to adjust the averaging. Press the
b
key
to set the value.
Select
[EXIT]
or press
a
to exit the menu.
NOTE: Increasing or decreasing the averaging setting has no effect on the data update rate.
Oximeter Messages
Oximeter messages are shown on the second line of the message area at the upper left of the display. If there are
several messages to be displayed, then they alternate, showing one message every second.
High Priority Alarm Messages
SpO2 > (x)
Indicates the SpO
2
measurement is equal to or higher than the high SpO
2
alarm limit. (x) is the high alarm limit
setting.
SpO2 < (x)
Indicates the SpO
2
measurement is equal to or lower than the low SpO
2
alarm limit. (x) is the low alarm limit
setting.
Pulse > (x)
Indicates the pulse rate measurement is equal to or higher than the high PULSE alarm limit. (x) is the high alarm
limit setting. (Only if
p
SOURCE = OXIMETER.)
Pulse < (x)
Indicates the pulse rate measurement is equal to or lower than the low PULSE alarm limit. (x) is the low alarm
limit setting. (Only if
p
SOURCE = OXIMETER.)
Lost Pulse
Indicates that the sensor no longer detects a pulse while a finger is inserted in the sensor, where a pulse was
previously detected.
1.
2.
3.
4.
Содержание SurgiVet V3404 Plus
Страница 2: ......
Страница 8: ...Table of Contents vi Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 12: ...Warranty and Service Information Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 20: ...Chapter 1 Introduction 1 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 50: ...Chapter 6 Oximetry 6 10 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 64: ...Chapter 10 Serial Output 10 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 66: ...Chapter 11 Analog Output 11 2 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 84: ...Appendix B Revision History B 2 Autocorr Plus Operation Manual This page is intentionally left blank ...
Страница 85: ......