Chapter 1: Introduction
Operation Manual SurgiVet
®
V3404 Plus
1-3
WARNING! The monitor is protected against damage from defibrillator, diathermy, and electrocautery
discharge.
WARNING! Where HF (diathermy) is used there is no danger of burning to the patient provided
recommended components are used. Rate meters may be temporarily affected.
WARNING! The vital signs monitor is suitable for use within the patient environment IEC 60950 approved
equipment must be placed outside of the patient environment. The patient environment is
defined as any volume in which intentional or unintentional contact can occur between the
patient and parts of the system or between the patient and other persons touching parts of
the system.
Figure 1.1: Patient Environment (Dimensions are not prescriptive)
WARNING! When connecting this monitor to any instrument, verify proper operation before clinical use.
Refer to the instrument’s user manual for full instructions. Accessory equipment connected
to the monitor’s data interface must be certified according to the respective IEC standards,
i.e., IEC 950 for data-processing equipment or IEC 601-1 for electromedical equipment. All
combinations of equipment must be in compliance with IEC 601-1-1 systems requirements.
Anyone connecting additional equipment to the signal input port or signal output port
configures a medical system, and, therefore, is responsible that the system complies with the
requirements of the system standard IEC 601-1-1.
WARNING! In the event that earth ground integrity is lost, the performance of this device and/or other
devices nearby may be affected due to excessive RF emissions.
WARNING! The monitor should be operated from its internal power source (if fitted) if the integrity of the
protective earth conductor is in doubt.
WARNING! Any monitor that has been dropped or damaged should be checked by a qualified service
personnel to insure proper operation prior to use.
WARNING! This monitor will not operate effectively on patients who are experiencing convulsions or
tremors.
WARNING! Patient safety can be compromised by the use of a power supply not supplied by Smiths
Medical PM, Inc. Use only the power supply included with your monitor, or one approved by
Smiths Medical PM, Inc.
WARNING! Ensure the device’s AC rating is correct for the AC voltage at your installation site before using
the monitor. The monitor’s AC rating is shown on the external power supply. If the rating
is not correct, do not use the monitor; contact Smiths Medical PM, Inc. Veterinary Division
service department for help.
WARNING! Ensure conductive parts including electrodes of the patient cable do not come into contact
with any conductive surfaces or earth parts.
Содержание SurgiVet V3404 Plus
Страница 2: ......
Страница 8: ...Table of Contents vi Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 12: ...Warranty and Service Information Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 20: ...Chapter 1 Introduction 1 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 50: ...Chapter 6 Oximetry 6 10 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 64: ...Chapter 10 Serial Output 10 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 66: ...Chapter 11 Analog Output 11 2 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 84: ...Appendix B Revision History B 2 Autocorr Plus Operation Manual This page is intentionally left blank ...
Страница 85: ......