Chapter 4: Alarms
Operation Manual SurgiVet
®
V3404 Plus
4-3
Adjusting or Viewing Alarm Limits
From the Main menu, select the
ALARMS
item.
Using the ARROW keys (
no
), select the alarm limit to be changed under the
LOW
or
HIGH
column.
Press the
b
key to highlight the value. Use the arrow keys to adjust the value. Press the
b
key to set the
value.
Select
[EXIT]
or press the a key to exit menu.
NOTE: Alarm limits are saved through power cycles, with the exception of the following note.
NOTE: If the operator changes the low SpO
2
alarm limit to a value less than 80, and a power down-power
up sequence takes place, a minimum value of 85 takes the place of the operator’s entered value.
NOTE: Alarms may be tested while the monitor is in use by setting alarm limits such that the measured
parameter is outside the alarm limits. Be sure to restore alarm limits to required settings after
testing.
Alarm Tones
The high priority alarm tone is a three-tone burst with a pause followed by a two-tone sound (beep beep-
beep pause beep-beep). The tone for low priority alarms is a two-tone sound with a pause (beep-beep, pause,
beep-beep).
The high and low priority alarm tones sound at the same volume.
The volume can be adjusted in the
SETUP/VOLUME
menu.
The volume can not be set to OFF.
All alarm tones can be silenced.
NOTE: Information signals, i.e. Low Battery, cannot be silenced.
Turning Alarm Tones On and Off
When the monitor is turned on, the alarm tones are silenced for two minutes. The Alarm Silence indicator (
)
on the EL Display flashes during the two minute time-out.
To silence the alarm tones indefinitely: Press and hold
B
for about three seconds; the Alarm Silence
indicator (
) lights steady.
NOTE: To comply with government requirements for patient monitoring, the indefinite alarm tone
silence feature may not be available in monitors shipped to your country.
NOTE: To comply with government requirements for patient monitoring, the indefinite alarm tone
silence feature may be turned off. For more information on turning the permanent silence feature
on and off, please see the Permanent Silence Section below.
To silence the alarm tones for two minutes: Momentarily press
B
; the Alarm Silence indicator (
)
flashes. If tones are already silenced, press
B
twice (the first press cancels alarm silenced; the second
press silences the alarms for two minutes).
To cancel either two minute or indefinite alarm silence and enable alarm tones: Momentarily press
B
;
the Alarm Silence indicator (
) lights steady and the slash is removed.
1.
2.
3.
4.
•
•
•
•
•
•
•
Содержание SurgiVet V3404 Plus
Страница 2: ......
Страница 8: ...Table of Contents vi Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 12: ...Warranty and Service Information Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 20: ...Chapter 1 Introduction 1 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 50: ...Chapter 6 Oximetry 6 10 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 64: ...Chapter 10 Serial Output 10 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 66: ...Chapter 11 Analog Output 11 2 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 84: ...Appendix B Revision History B 2 Autocorr Plus Operation Manual This page is intentionally left blank ...
Страница 85: ......