8
Note This procedure will cover only the specific steps of freehand targeting of intramedullary locking
holes using the TRIGEN™ SURESHOT™ Distal Targeting System. For the full surgical procedure, please
refer to the specific surgical technique for the TRIGEN IM Nail System being implanted.
Surgical procedure – OR preparation
Trauma Interface setup
After the sterile areas have been established,
place the Trauma Interface (7169-2802) in the
desired non-sterile location and turn on the
power switch.
Tip If the Trauma Interface does not power on,
make sure the switch is in the “on” position.
Note No other electrical devices should be
placed near the Trauma Interface. See the
“Guidance and Manufacturer’s Declaration –
Separation Distances” table at the end of
this document.
When start-up screen disappears before the
display prompts for tool connections, a warning
screen will appear: Do not operate the TRIGEN
SURESHOT Targeter within 200mm of an installed
pacemaker.
The magnetic field produced by the Targeter may
interfere with the operation of the pacemaker.