20

4

 

A software window will appear informing 
the user if the TRIGEN™ SURESHOT™ Targeter and 
Probe combination is within the  
predefined accuracy parameters (“Pass”  
or “Fail” message).

5

 

If the field accuracy check fails, check the 
“Troubleshooting” section of this document for 
possible solutions.

Software upgrades 
Software upgrades are performed using a  
Smith & Nephew, Inc. supplied USB memory 
stick and a Work Instruction document. The Work 
Instruction document should be filled out and 
returned to Smith & Nephew Customer Service 
upon completion.

Cleaning and disinfection 
The TRIGEN SURESHOT Trauma Interface is used 
in the non-sterile area of the operating theatre 
and is cleaned and disinfected using commercial 
cleaning and disinfectant products such as a mild 
detergent and water or a bactericidal cleaning 
solution such as 70% isopropyl alcohol.
Care must be taken not to allow any liquid to 
pass into any electrical connections or the  
interior of the unit.  Let the surfaces dry  
thoroughly before plugging in the unit. DO NOT 
steam sterilize the system. DO NOT submerge 
the system for any reason.
All other reusable instruments shall be cleaned 
and sterilized according to the 

Cleaning and 

Sterilization Instructions

 (Smith & Nephew 

document 7138-1339).

Note Observe the manufacturer’s instructions for 
dilution, exposure time, etc.
Note DO NOT sterilize with ethylene oxide gas or 
steam sterilize the Trauma Interface.
Note Probes are NOT reusable.

Содержание TRIGEN SURESHOT

Страница 1: ...User Manual V2 1...

Страница 2: ......

Страница 3: ...creen operation 17 Shutting down the system storage and transport 18 Maintenance 19 Service 22 Troubleshooting 23 Product label and transportation case label 25 Technical specifications 26 Parts and a...

Страница 4: ...SHOT Targeter within 200mm of an installed pacemaker The magnetic field produced by the Targeter may interfere with the operation of the pacemaker Intended use The TRIGEN SURESHOT Distal Targeting Sys...

Страница 5: ...tive Made in USA 81068350 Rev C Trauma Interface 71692802 Patents Pending T3 15A 250V 5 x 20mm 100 240V 50 60 Hz 2 2A 2X Kommunikationsschnittstelle f r Trauma Interfaz de trauma Interface trauma Inte...

Страница 6: ...mitted to install or uninstall software Any new software must be installed by the manufacturer or by authorized personnel It is only allowed to connect equipment to the interface and power supply conn...

Страница 7: ...running Note Danger of damage and tipping over Tip Place the unit on a firm level surface capable of holding at least 10kg 22 lbs Note To avoid the risk of electric shock this equipment must only be...

Страница 8: ...untry specific Front panel layout 1 Probe sensor ports 2 Targeter port 3 Touch screen interface Rear panel layout 1 Power switch 2 Power connection 3 USB connections service only 4 VGA video output 10...

Страница 9: ...w Inc Note This device is provided non sterile and must be cleaned and sterilized per Cleaning and Sterilization Smith Nephew document 7138 1339 prior to use Note The probe will be used as an intramed...

Страница 10: ...802 in the desired non sterile location and turn on the power switch Tip If the Trauma Interface does not power on make sure the switch is in the on position Note No other electrical devices should be...

Страница 11: ...er to ensure proper accuracy Tip When oriented as shown the connector should assemble easily Do not force the connector into the port Note If the Targeter is properly connected to the system and the a...

Страница 12: ...T META NAIL Semi extended Drill Guide Probe 7169 2803 TRIGEN SURESHOT Percutaneous TAN FAN Drill Guide Probe 7169 2815 TRIGEN SURESHOT Humeral Nail Drill Guide Probe 7169 1152 Red probe Use only with...

Страница 13: ...ase the probe has to be exchanged Tip It is possible at any time to disconnect and reconnect tools when the application is running The display will show a screen reporting the missing instrument Confi...

Страница 14: ...eve to the Targeter Tip The drill sleeve 7169 2804 or 7169 2805 can be loosened from the Targeter using the slot in the TriGEN Hammer 7167 4082 Tip The drill sleeve 7169 1154 can be loosened from the...

Страница 15: ...curacy the accuracy check has to be performed directly in the operative field Place the IM nail with inserted probe directly beside the patient s limb Position the Targeter over the locking holes and...

Страница 16: ...there is metal or electrical interference the green and red Targeter circles on the navigation screen may become unstable and or a warning message will be displayed If the interference is excessive th...

Страница 17: ...tic TRIGEN SURESHOT drill bit or long gold non magnetic drill bit Tip After measuring the length on the drill bit prior to drilling the far cortex add the appropriate screw length in order to account...

Страница 18: ...m Gold Drill Sleeve 7163 1152 or 7175 1128 down to bone over the hexdriver The screw head will be flush with bone when the groove on the hexdriver is even with the end of the 9 0mm Gold Drill Sleeve T...

Страница 19: ...interlocking holes The white lines displayed on either side of the IM nail can be used for targeting blocking screws These lines are located 2 5mm from the side of the IM nail for all IM nails 10mm an...

Страница 20: ...case Consider the temperature and humidity range for storage and transportation on the transportation case Note Shut down the Trauma Interface using the on screen Shut Down option and wait for the sy...

Страница 21: ...components are working correctly prior to their use on a patient Field accuracy check steps 1 Attach TRIGEN SURESHOT Field Accuracy Gauge 7169 2808 to TRIGEN SURESHOT Targeter The knob on the Field Ac...

Страница 22: ...sterile area of the operating theatre and is cleaned and disinfected using commercial cleaning and disinfectant products such as a mild detergent and water or a bactericidal cleaning solution such as...

Страница 23: ...echnical Specifications section for replacement fuse type Note Disconnect the power cord before exchanging any fuse Note To avoid fire hazard use only fuses of the correct type voltage rating and curr...

Страница 24: ...nt contains electronic printed circuit assemblies At the end of the useful life of the equipment it should be disposed of in accordance with any applicable national or institutional related policy rel...

Страница 25: ...ter Unplug Targeter wait 10 seconds plug back in Damaged Targeter Replace Targeter with new unit Probe not recognized Error reading data from probe Unplug probe wait 10 seconds plug back in Damaged pr...

Страница 26: ...ter Over tightening of drill sleeve Use the Slotted Hammer from the instrument tray as a wrench to unscrew the drill sleeve counter clockwise from the Targeter The 4 7mm 4 0mm step drill will not fit...

Страница 27: ...r Trauma Interfaz de trauma Interface trauma Interfaccia Trauma Traumagr nssnitt Trauma interface Interface para Traumatismo Traumeinterface Traumegrensesnitt 0086 Smith Nephew Inc Memphis TN 38116 U...

Страница 28: ...system meets IEC standards eg IEC 60601 All configurations comply with standard IEC 60601 1 1 Any person connecting equipment to the system is responsible for the configuration and must ensure that i...

Страница 29: ...rill Guide Probe 7169 2815 TRIGEN SURESHOT Percutaneous TAN FAN Drill Guide Probe 7169 2816 TRIGEN SURESHOT TAN Anteversion Locking Guide 7169 2817 Trauma Interface Case 7169 2819 TRIGEN SURESHOT TAN...

Страница 30: ...d or adverse operation Note The equipment or system shall not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the equipment or system shall be tested to ver...

Страница 31: ...e TRIGEN SURESHOT Distal Targeting System uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipm...

Страница 32: ...ty should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60...

Страница 33: ...s determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol...

Страница 34: ...on equipment Separation distance according to frequency of transmitter m Rated maximum output power of transmitter W 150 kHz to 80 MHz d 1 2 80 MHz to 800 MHz d 1 2 800 MHz to 2 5 GHz d 2 3 0 01 0 12...

Страница 35: ...able for refurbishment may result in additional charges regardless of warranty status All warranties apply to the original buyer only In no event shall Smith Nephew be liable for and anticipated profi...

Страница 36: ...horization RA number and returned postage paid to the appropriate Smith Nephew Customer Service Center Smith Nephew Customer Service or your local authorized representative can provide shipping inform...

Страница 37: ...35 Notes...

Страница 38: ...36 Notes...

Страница 39: ......

Страница 40: ...208 135 Orthopaedics Smith Nephew Inc 7135 Goodlett Farms Pkwy Cordova TN 38016 USA Telephone 1 901 396 2121 Information 1 800 821 5700 Orders and Inquiries 1 800 238 7538 www smith nephew com 2011 S...

Отзывы: