Probe connection
Connect the probe to either of the probe sensor
ports on the Trauma Interface.
Tip When oriented as shown, the connector
should assemble easily. Do not force the
connector into the port, simply try rotating the
connector until the keys are oriented in the
12 o’clock position.
Note If the probe is properly connected to the
system and the application reports “Probe not
found” for more than 10 seconds, the probe may
be damaged or defective. In this case, the probe
has to be exchanged.
Tip It is possible at any time to disconnect and
reconnect tools when the application is running.
The display will show a screen reporting the
missing instrument.
Confirm that the tool connection has
been verified.
Drill sleeve selection
Select the length of the drill sleeve (7169-2804 or
7169-2805) that will be used.
Tip A different sleeve can be selected at any
time during the procedure by choosing the drill
sleeve option from the drop down menu after the
implant has been selected. The selected sleeve
will be noted on the navigation screen.
Note For Humeral Nails, there is only one option
provided for the drill sleeve (7169-1154)