7
Applied parts for targeting
Note The Targeter will be operated within the
sterile field and may have contact with the skin
of the patient. The drill sleeve inserts will be used
in the incision and have direct bone contact.
Note Verify that the Targeter housing is not
damaged (holes, tears, cracks). If the housing
or the connector is damaged, the Targeter is no
longer safe to use.
Note If the Targeter is not recognized after
its connection to the system, the Targeter is
defective and must be exchanged. (See also
instrument connection).
Note Broken or damaged instruments must
be exchanged immediately and sent back to
Smith&Nephew, Inc.
Note This device is provided non-sterile and
must be cleaned and sterilized per
Cleaning
and Sterilization
(Smith & Nephew document
7138-1339) prior to use.
Note The probe will be used as an intramedullary
tool inside the nail placed in the patient’s bone.
Note If the probe is not recognized after its
connection to the system, the probe is
defective and must be exchanged. (See also
instrument connection).
Note Broken or damaged instruments must
be exchanged immediately and sent back to
Smith & Nephew, Inc.
Note This device is provided sterile by ethylene
oxide gas and is single use.
Note In case the packaging is compromised,
please replace with new sterile packaged probe.
TRIGEN™ SURESHOT™ Targeter
(Body Applied Part Type BF)
TRIGEN SURESHOT Probes
(Body Applied Part Type BF)
TRIGEN SURESHOT META-NAIL™ Standard Drill Guide Probe