11
The ORTHOMATCH™ femoral sizing guide allows for
external rotation to be set from 0-9º based on surgeon
preference and patient anatomy. Rotational alignment
may be checked by aligning the A/P axis (Whiteside’s
Line) with the vertical marks on the sizing guide or by
ensuring that the “EPI” lines on the face of the guide
are parallel with the epicondylar axis. Once rotation
is set the guide can be locked by turning the quick
connect handle.
The sizing guide can be used for either posterior
referencing or anteriorly referencing by utilizing either
the “Ant Ref” or “Post Ref” holes. These holes correlate
with similar holes on the A/P resection blocks.
The femoral size is indicated on either side of the
femoral housing of the sizer. In addition, the femoral
stylus is marked with sizing correlating with the femoral
flange length / anterior sawblade exit point.
Femoral positioning and sizing
Stylus marked to indicate anterior femoral
flange length/sawblade exit point
Quick connection handle attaches to sizing
guide for application to resected distal femur
Size reading
Posterior
referencing
holes
Locking knob
Posterior
paddles
Femoral external
rotation guide
(0-9 degrees)
Fixation holes
Anterior referencing
holes
Trans-
Epicondylar
axis (TEA)
A/P axis
(Whiteside’s line)
Содержание Syncera Anthem