8
4.
Align the extramedullary tibial alignment guide parallel
to the tibial axis in the coronal and sagittal planes.
Rotate the assembly to the medial one-third of the tibial
tubercle and impact the anterior spike of the up rod
(Figure 10 and 10a).
Note: 3° of posterior slope is built into the cut block
connector and 4° of slope is built into the articular
insert.
5.
Attach the tibial stylus to the universal cutting block by
inserting the foot of the tibial stylus into the cutting slot
of the universal cutting block.
6.
Adjust the resection level by lowering the universal
cutting block until the tibial stylus touches the low
point on the less affected side of the tibia. The tibial
stylus can be adjusted for 2mm (affected side) or
9mm (unaffected side) tibial resection by reversing its
position in the cutting slot of the universal cutting block
(Figure 11).
7.
Pin the universal cutting block to the tibia by inserting
pins first through the central holes; then the medial
hole (Figure 11a).
Figure 10a
Figure 11
Figure 11a