Operation
06/18/04
8-9
8.6
Detail of setting mode level
The each setting mode level is explained below in detail.
8.6.1 Setting mode, Level 1
1.Target temp. (no indication on display)
Setting range : 4.0 to 45.0deg.C
Min. increment : 0.5deg.C(manual setting)
Indicated content : [SV] : Target temp.(Ex. : 30.0deg.C)
Function : Sets target temp.
2.Set Gradient Temp.
Setting range : 10 to 60deg.C/h
Min. increment : 5deg.C
Indicated content : Target gradient temp.(Ex. : 60deg.C/h)
Function : Sets target gradient temp.
3.Control Operation
Setting range : 0,1,2
Indicated content : Number and description of control
operation mode.(Ex. : 1:Normal)
Function : Selects control operation mode among them
below
0 : Pump Stop(IDC setting)
1 : Normal(normal control operation)
2 : AT(auto tuning)
4.Safety Default Set Temp.
Setting range : 4.0 to 45.0deg.C
Min. increment : 0.5deg.C(manual setting)
Indicated content : Safety default temp.(Ex. : 30deg.C)
Function : Sets temp. which causes SWT time out to
arise(i.e. no communication command is input
within 10sec).
THERMO-CON
P V
2 5 . 0 ° C # 1
S V < 3 0 . 0 ° C
MO D E
THERMO-CON
S e t G r a d i e n t
T e m p .
MO D E < 6 0 ° C / h
THERMO-CON
C o n t r o l
Op e r a t i o n
MO D E < 1 : N o r m a l
THERMO-CON
S a f e t y D e f a u l t
S e t T e mp .
MO D E < 3 0 . 0 ° C