Specifications
06/18/04
6-1
6 Specifications
The specifications of the Thermo-con is as shown below.
Table6-1 Specifications
Item
Spec.
Set Temp. range
4.0 to 45.0deg.C(min. increment : manual setting 0.5deg.C
Communication setting 0.1deg.C)
No dew condensation
Set gradient temp.
10 to 60deg.C/h (min. increment : 5deg.C/h)
Temp. measured range
-9.9 to 80.0deg.C(min. increment : 0.1deg.C)
Ambient environment
Temp. :10 to 40deg.C(No dew condensation and icing)
Max. gradient temp. : 20deg.C/h
Humidity : 10 to 80%
Max. gradient humidity : 30%/h
Atmospheric pressure : 106 to 70kPa(sea level altitude-100 to 3000m)
Max. gradient atmospheric pressure : 1.8kPa/h
Environment : No corrosive gas, solvent such as thinner and flammable gas
Storage environment
Temp. : -40 to 70deg.C(No dew condensation and icing)
Max. gradient temp. : 20deg.C/h
Humidity : 5 to 95%
Max. gradient humidity : 30%/h
Atmospheric pressure : 106 to 12kPa
Max. gradient atmospheric pressure : 76kPa/h
Environment : No corrosive gas, solvent such as thinner and flammable gas
Storage period (before
initial operation)
15 weeks with no damage
Accuracy related to temp.
Setting accuracy : ±0.2deg.C at max.
Temp. drift : ±0.2deg.C at max.(Ambient 10 to 40deg.C)
Control accuracy : ±0.05deg.C at max.
Cooling capacity
(Distilled water)
Approx.105W
Conditions : Set temp.- Ambient temp =-15deg.C, Circulating liquid flow rate
2.6L/min, Atmospheric pressure 101kPa
Circulating liquid
Circulating liquid type : Distilled water, Propylene glycol 25% (Reference GEMS
spec #2352900PSP; chapter10)
Pump capacity : 1.25L/min at 0.0482MPa or more(by distilled water)
Reservoir capacity : Approx.0.8L
Max. pressure : 0.09MPa(Gauge)
Fitting : for circulating liquid : PLCD16006(IN/OUT)
for drain : Rc1/4(with plug)
Material for wetted part
SUS303, SUS304, EPDM, Acetal resin, Ceramic, PPS glass 30%, Carbon
PE, Polyurethane