SLS Lab Basics 12 Place Microcentrifuge User Manual
5
This section describes the instructions that you should abide when installing the centrifuge to ensure your
safety and the optimum performance. Before moving the centrifuge, the rotor must be removed.
4. Installation
•
Improper power supply may damage centrifuge
•
Make sure the power source conforms to the required power supply before connecting
WARNING:
4.1 Location
(1) Place the centrifuge on a firm, flat and level table, ensure the four feet of this centrifuge stand on the
table firmly. Avoid installing on the slippery surface or surface prone to vibration.
(2) Ideal ambient temperature is 20ºC ±5ºC, avoid placing the centrifuge in direct sunlight if temperature
exceeds 30ºC.
(3) Keep clear of the centrifuge at least 10cm on both sides and at least 30cm behind it to guarantee the
cooling efficiency.
(4) Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
4.2 Connection of the power cord and grounding
A minimum 10A outlet providing a sufficient ground is required, and this must meet with local safety
requirements.
•
To avoid electrical shocks, ensure your hands are dry when touching the power cord
•
This centrifuge must be grounded properly
WARNING:
SLS Lab Basics 12 Place Microcentrifuge User Manual