SLS Lab Basics 12 Place Microcentrifuge User Manual

10

High Speed Mini Micro-Centrifuge 

 

 

10 

 

 

Figure 8-2 the rotor installation 

 

 

Load the rotor to the shaft to ensure rotor is in position until it connected well with the shaft. 

 

Rotate the rotor slightly by your fingers to check, if the rotor vibrates, if so attach the rotor again. 

 

Hold on the rotor with one hand, tighten clockwise the nut with the other hand, and make sure tighten 
firmly. 

 

Put the rotor lid to  the rotor, press the fastener down and make the fastener attach  the  locking  nut 
tightly. 

 

Close the door and start running. 

 

To release the rotor, firstly  pull  the  fastener  up to remove the lid, and then  turn the locking  nut 
anti-clockwise. 

 

  CAUTION 

 

Check the rotor is firmly tightened before running. 

 

8.1.2

 

Set the operation parameters 

Press the 

  button to select required parameters. The parameter can be modified when the parameter is 

flashing. Clockwise rotate the parameter button

to increase parameter value; counter-clockwise rotate the 

parameter  button 

  to  decrease  parameter  value.  Parameter  button

rotate faster, parameter  value 

increase faster. The minimum speed increment is 100 rpm, the minimum time increment is 1 second. 
 

1

Set the speed

 

 

Press the select button

  until the speed rpm is displayed. 

 

When the speed button is selected, the speed symbol will flash the speed value. 

 

The minimum speed value you can set 500rpm, the minimum increment is 100rpm. 

 

Rotate program button clockwise 

to increase speed value, rotate the program button anti-clockwise 

to decrease speed value. 

 

You can speed-up set the speed value by rotating program button

  quickly. 

 

There is a circulating function to increase/decrease  the speed values.  Rotate  the program  button 

Fastener 

Figure 8.2   The rotor installation

•  Load the rotor to the shaft to ensure rotor is in position until it is connected well with the shaft
•  Rotate the rotor slightly by your fingers to check, if the rotor vibrates, if so attach the rotor again
•  Hold on the rotor with one hand, tighten clockwise the nut with the other hand and make sure to 

tighten firmly

•  Put the rotor lid to the rotor, press the fastener down and make the fastener attach the locking nut tightly
•  Close the door and start running
•  To release the rotor, firstly pull the fastener up to remove the lid, and then turn the locking nut
       anti clockwise

CAUTION:

• 

Check the rotor is firmly tightened before running

8.1.2  Set the operation parameters

Press the         button to select required parameters. The parameter can be modified when the parameter 
is flashing. Clockwise rotate the parameter button              to increase parameter value, counter clockwise 
rotation of the parameter button will decrease the parameter value. When the Parameter button is rotated 
faster – the parameter value increases faster. The minimum speed increment is 100 rpm, the minimum 
time increment is 1 second.

Select

(1) 

 

Set the speed

•  Press the select button          until the speed rpm is displayed
•  When the speed button is selected, the speed symbol will flash the speed value
•  The minimum speed value you can set 500rpm, the minimum increment is 100rpm
•  Rotate program button             clockwise to increase speed value, rotate the program button 
       anti-clockwise to decrease speed value
•  You can speed up set the speed value by rotating program button quickly

Select

Содержание D2-236

Страница 1: ...eland NI sales scientific labs com Republic of Ireland ROI sales scientific labs com High Speed Mini Centrifuge Before using centrifuge please carefully read this user manual for its efficient operati...

Страница 2: ...nto tubes 8 7 3 Keep the tubes balanced 8 7 4 Inspect the rotor 8 7 5 Symmetrically load centrifuge tubes in rotor 8 8 Operation 9 8 1 Normal Operation 9 8 2 RCF Operation 12 8 3 Pulse Operation 12 9...

Страница 3: ...bserved could result in personal injury or possible death CAUTION Possible damage to instrument Caution notes indicate any condition or practice which if not strictly observed or remedied could result...

Страница 4: ...vel table Make sure the centrifuge is horizontal before running Make sure the angle between the door and cover is greater than 70 degrees when the door is open Be careful not put your fingers or hands...

Страница 5: ...onstruction in accordance with the following safety standards EN 61010 1 Construction in accordance with the following EMC standards EN 61326 1 FCC Part 15 Subpart B ICES 001 Associated EU guidelines...

Страница 6: ...ment The centrifuge generates uses and can radiate radio frequency energy and if not installed and used in accordance with the user manual may cause harmful interference to radio communications Operat...

Страница 7: ...d installing on the slippery surface or surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if temperature exceeds 30 C 3 Keep clear of t...

Страница 8: ...Door Lock Hook Foot Operation Panel Door View Window Air Vents Power Inlet Power Switch Figure 5 1 Front view of centrifuge Figure 5 2 Rear view of centrifuge 6 Door release hole SLS Lab Basics 12 Pla...

Страница 9: ...m running time is 12 minutes When speed symbol is rotating indicating the centrifuge is running Time display symbol displays the ratio of working to time setting The total time setting is divided into...

Страница 10: ...nder unbalanced condition even though the allowable imbalance is not exceeded 7 4 Inspect the rotor Check the rotor for corrosion or scratches before using 7 5 Symmetrically load centrifuge tubes in r...

Страница 11: ...tton shift between speed and RCF display 4 Open lock button Press the button to open the door The button is not available when the centrifuge is running 5 Start Stop button Press the button to start r...

Страница 12: ...e speed value rotate the program button anti clockwise to decrease speed value Fastener Figure 8 2 The rotor installation Load the rotor to the shaft to ensure rotor is in position until it is connect...

Страница 13: ...ter button to modify values Release the button after 5 seconds and the centrifuge will return to normal operation mode and run according to the new value If the set time value has been modified the op...

Страница 14: ...hten the rotor lid and make sure it is in secured position and then close the door 2 The centrifuge gets into preparation mode and displays last running values The values can be reset 3 Press knob and...

Страница 15: ...coat of silicon grease to the rotor when it is completely dry The rotor should be regular maintained we recommend cleaning it every 3 months to ensure tube and rotor holes keep clean and then apply a...

Страница 16: ...ter the cause removed 10 2 How to open the door 10 2 1 In the case of power on CAUTION The door can be opened while the power is on and the rotor stops rotating 1 Turn on the POWER switch the door loc...

Страница 17: ...tube 0 2ml PCR A05P2 15000 11700 Locking nut Fastener Lid Rotor block Spindle hole Tube Locking Nut Fastener Lid Rotor Block Spindle Hole Tube Figure 11 1 The rotor profile 11 1 2 Available rotors and...

Страница 18: ...iling 15 to 30 minutes O O O Ultraviolet sterilization 200 300nm X X X 11 1 3 Notice The centrifuge rotor can separate samples with density lower than 2 0g ml If the separated samples density is over...

Страница 19: ...aced sideways they may deform into an oval shape due to gravity 2 Remove screw nuts and inner covers to prevent from deformation or rupture 3 Wait until the sterilising chamber cools down to the room...

Страница 20: ...igue damage We do not guarantee this centrifuge and the rotor under the following conditions even if within the guarantee period 1 Failures caused by incorrect installation 2 Failures caused by rough...

Страница 21: ...dependent channel partner to a vast range of liquid handling solutions we offer pipette servicing solutions for an extensive range of brands On Site Repairs For an agreed fee the SLS Service Centre wi...

Отзывы: