8. Operation

CAUTION:

• 

Do not push or lean against the centrifuge while it is running

• 

Do not run the centrifuge when fragments or sample solutions are left in the centrifuge chamber

• 

Always keep the centrifugal chamber clean

• 

If the centrifuge makes strange noise during operation, stop it immediately and contact the SLS 
Service Centre. Notify them of the warning code if displayed

8.1 Normal Operation

Turn on the power switch, centrifuge will display the running interface last time after passing the self-
diagnostic checks, see figure 8.1 below:

High Speed Mini Micro-Centrifuge 

 

 

6. Operation panel                         

 

Figure 6-1   Operation Panel 

 

Item 

Symbol 

Name 

Function 

 

Select button 

Press the button  to choose the program  which you want to 
modify. 

2

 

 

Pulse button 

The speed can be accelerated and held at the speed when 
pressing Pulse on. 

 

Parameter button 

Clockwise rotate  to  increase  program  values.  Rotate 
anti-clockwise to decrease parameter values. 
Press the button, shift between speed and RCF display.   

 

Open/ lock button   

Press the button to open the door The button is not available 
when the centrifuge is running. 

5

 

 

Start/ Stop button 

Press the button to start running. The centrifuge will brake to 
stop running if pressed during centrifugation. 

 

 
 
 
 
 
 
 
 

 

Figure 6-2 the main interface 

 

Main interface is as figure 6-2. The speed is set to be 15000 rpm, running time is 12 minutes. When speed 

symbol is rotating, indicating the centrifuge  is running.  Time display symbol   display the ratio  of 
working to time setting. The total time setting is divided into 10 sections. 

 

Speed area     Lock status     Time area

 

 

 

150          12

 

min

 

 

×

×

100rpm 

                      

 

 

Figure 8.1   The last running interface

•  Speed: 15000 rpm Running time: 12 minutes
•  The door lock is released

8.1.1 Rotor loading and removal

CAUTION:

• 

Attach the rotor to the rotor shaft. Ensure the rotor is in position and connected with the shaft,

      tightening the locking nut to secure the rotor with shaft, to prevent the rotor damaging the centrifuge

• 

Ensure the rotor lid is firmly tightened to the rotor

9

SLS Lab Basics 12 Place Microcentrifuge User Manual

Содержание D2-236

Страница 1: ...eland NI sales scientific labs com Republic of Ireland ROI sales scientific labs com High Speed Mini Centrifuge Before using centrifuge please carefully read this user manual for its efficient operati...

Страница 2: ...nto tubes 8 7 3 Keep the tubes balanced 8 7 4 Inspect the rotor 8 7 5 Symmetrically load centrifuge tubes in rotor 8 8 Operation 9 8 1 Normal Operation 9 8 2 RCF Operation 12 8 3 Pulse Operation 12 9...

Страница 3: ...bserved could result in personal injury or possible death CAUTION Possible damage to instrument Caution notes indicate any condition or practice which if not strictly observed or remedied could result...

Страница 4: ...vel table Make sure the centrifuge is horizontal before running Make sure the angle between the door and cover is greater than 70 degrees when the door is open Be careful not put your fingers or hands...

Страница 5: ...onstruction in accordance with the following safety standards EN 61010 1 Construction in accordance with the following EMC standards EN 61326 1 FCC Part 15 Subpart B ICES 001 Associated EU guidelines...

Страница 6: ...ment The centrifuge generates uses and can radiate radio frequency energy and if not installed and used in accordance with the user manual may cause harmful interference to radio communications Operat...

Страница 7: ...d installing on the slippery surface or surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if temperature exceeds 30 C 3 Keep clear of t...

Страница 8: ...Door Lock Hook Foot Operation Panel Door View Window Air Vents Power Inlet Power Switch Figure 5 1 Front view of centrifuge Figure 5 2 Rear view of centrifuge 6 Door release hole SLS Lab Basics 12 Pla...

Страница 9: ...m running time is 12 minutes When speed symbol is rotating indicating the centrifuge is running Time display symbol displays the ratio of working to time setting The total time setting is divided into...

Страница 10: ...nder unbalanced condition even though the allowable imbalance is not exceeded 7 4 Inspect the rotor Check the rotor for corrosion or scratches before using 7 5 Symmetrically load centrifuge tubes in r...

Страница 11: ...tton shift between speed and RCF display 4 Open lock button Press the button to open the door The button is not available when the centrifuge is running 5 Start Stop button Press the button to start r...

Страница 12: ...e speed value rotate the program button anti clockwise to decrease speed value Fastener Figure 8 2 The rotor installation Load the rotor to the shaft to ensure rotor is in position until it is connect...

Страница 13: ...ter button to modify values Release the button after 5 seconds and the centrifuge will return to normal operation mode and run according to the new value If the set time value has been modified the op...

Страница 14: ...hten the rotor lid and make sure it is in secured position and then close the door 2 The centrifuge gets into preparation mode and displays last running values The values can be reset 3 Press knob and...

Страница 15: ...coat of silicon grease to the rotor when it is completely dry The rotor should be regular maintained we recommend cleaning it every 3 months to ensure tube and rotor holes keep clean and then apply a...

Страница 16: ...ter the cause removed 10 2 How to open the door 10 2 1 In the case of power on CAUTION The door can be opened while the power is on and the rotor stops rotating 1 Turn on the POWER switch the door loc...

Страница 17: ...tube 0 2ml PCR A05P2 15000 11700 Locking nut Fastener Lid Rotor block Spindle hole Tube Locking Nut Fastener Lid Rotor Block Spindle Hole Tube Figure 11 1 The rotor profile 11 1 2 Available rotors and...

Страница 18: ...iling 15 to 30 minutes O O O Ultraviolet sterilization 200 300nm X X X 11 1 3 Notice The centrifuge rotor can separate samples with density lower than 2 0g ml If the separated samples density is over...

Страница 19: ...aced sideways they may deform into an oval shape due to gravity 2 Remove screw nuts and inner covers to prevent from deformation or rupture 3 Wait until the sterilising chamber cools down to the room...

Страница 20: ...igue damage We do not guarantee this centrifuge and the rotor under the following conditions even if within the guarantee period 1 Failures caused by incorrect installation 2 Failures caused by rough...

Страница 21: ...dependent channel partner to a vast range of liquid handling solutions we offer pipette servicing solutions for an extensive range of brands On Site Repairs For an agreed fee the SLS Service Centre wi...

Отзывы: