SLS Lab Basics 12 Place Microcentrifuge User Manual
Rotor lid loose
Tighten the rotor lid firmly
E–02 Door fault
The door opened in running
The Run/Stop button is pressed
while the door opening
Close the door immediately
Close the door and then start
to operate.
E-06
Set wrong speed
The setting speed exceed the
allowable range
Modify the speed value
E-10–86
Read the service manual
Contact the SLS Service Centre
Alarm c
ode
appear
ed on the
time displa
y scr
een
Table 10.1 1 Frequent problems and solutions
• Warning code E-1–9 are related to wrong operating. You can continue running the centrifuge after the
cause removed
10.2 How to open the door
10.2.1 In the case of power on
CAUTION:
•
The door can be opened while the power is on and the rotor stops rotating
(1) Turn on the POWER switch the door lock will release automatically.
(2) The door lock will release automatically once the operation finished.
(3) It is available to release the door by pressing button once the rotor stops.
Door
10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available to be opened manually.
(1) Ensure if the rotor has stopped rotating.
• Listen carefully to ensure no rotating sound can be heard
(2) Insert a screw driver into the hole to open the door.
• Hole is located on the top right side of the unit
• Insert a screw driver into the hole and push forward to release the door