Page 12
210-PMTXM Issue L
www.skcltd.com
Getting Started
1) Charging the Battery Pack
The Pocket Pump MTX series pumps incorporate an intelligent charging function that monitors the battery pack
during charging and automatically switches to a trickle charging rate when the battery pack is fully charged to prevent
damage due to over-charging. A full charge will take up to 6 hours dependent on the charge state of the battery
pack at the start of charging. The charging function also includes a feature that enables full charge rate only when
the battery pack temperature is within the range +5 to +38°C. When the temperature is outside this range only trickle
charging rate is enabled.
Pocket Pump battery packs can also be charged when not fitted to the pump. In this case charging occurs only at the
trickle charging rate and a full charge will take up to 24 hours.
Use only the correct SKC chargers (Part Nos. 223-229A - single charger, 223-107A - five station charger). The
chargers are supplied with mains input plugs suitable for use in the UK, Europe, USA and Australia / New Zealand.
Prior to first use the battery pack should be fully charged, ideally overnight. It will take 2 – 3 full charge / discharge
cycles for the battery pack to achieve optimum capacity. If a new battery pack repeatedly exhibits a charging fault as
detailed below it is recommended to condition the battery pack by running the pump until it shuts down due to low
battery, then remove the battery pack from the pump and charge for 24 hours.
To charge the battery pack select the correct mains input plug and fit it to the charger. Plug the charger into the
electrical mains supply and switch on the power. Plug the charger output jack plug into the mating socket at the base
of the pump. The LED indicator on the charger will illuminate to indicate that the battery pack is connected.
During charging all three bars of the battery status icon on the pump LCD screen will flash. At the end of charging the
battery status icon will be on (not flashing) and showing three bars to indicate full charge.
If the battery icon shows one flashing bar at the end of charging, this indicates a charging fault. It is recommended to
disconnect the charger, wait for the battery pack temperature to stabilise to the ambient temperature then reconnect
the charger. If the charging fault condition re-occurs this would indicate a fault with the pump, battery pack or charger.
Contact SKC Limited customer services for further advice.
Содержание 210-1002MTX
Страница 9: ...www skcltd com 210 PMTXM Issue L Page 7 Blank Page ...
Страница 32: ...Page 30 210 PMTXM Issue L www skcltd com Pocket Pump MTX Components 1 3 1 5 6 1 2 1 7 4 1 4 1 4 1 4 1 ...
Страница 38: ...Page 36 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 40: ...Page 38 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 42: ...Page 40 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 44: ...Page 42 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 46: ...Page 44 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 48: ...Page 46 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 50: ...Page 48 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 53: ...www skcltd com 210 PMTXM Issue L Page 51 Blank Page ...