Page 18
210-PMTXM Issue L
www.skcltd.com
Operating the Pocket Pump in Constant Flow Mode
2) Setting and Calibrating the Pump Flow Rate
Ensure that the pump is set to Constant Flow mode and that the battery is fully charged.
Set up the sample train as specified in the sampling method and connect to the pump air inlet port. Connect
a suitable calibrated flowmeter (such as a chek-mate or rotameter) to the inlet of the sample train.
Press [ ] to start the pump running.
Press . The LCD screen will show the flow rate setting screen with ‘
SET
’
flashing.
Use the and buttons to set the required flow rate in millilitres per minute on the
LCD screen.
When the required flow rate is set press the button to display the flow rate adjust
screen, with ‘
ADJ
’ flashing.
Observe the indicated flow rate reading of the flowmeter. Use the and buttons to
adjust the pump flow rate up or down, to precisely calibrate the required flow rate.
Содержание 210-1002MTX
Страница 9: ...www skcltd com 210 PMTXM Issue L Page 7 Blank Page ...
Страница 32: ...Page 30 210 PMTXM Issue L www skcltd com Pocket Pump MTX Components 1 3 1 5 6 1 2 1 7 4 1 4 1 4 1 4 1 ...
Страница 38: ...Page 36 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 40: ...Page 38 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 42: ...Page 40 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 44: ...Page 42 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 46: ...Page 44 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 48: ...Page 46 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 50: ...Page 48 210 PMTXM Issue L www skcltd com Intrinsic Safety Certification ...
Страница 53: ...www skcltd com 210 PMTXM Issue L Page 51 Blank Page ...