Troubleshooting
Possible Fault
Corrective Action
Pump stops due to low battery
before the end of the required
sample period
• Battery pack not fully charged before starting sample run. Ensure
battery pack is fully charged before starting a sample run.
•
Battery pack has reduced capacity as it nears end of life. Replace the
battery pack.
•
Sample tube back pressure too high. Try a lower flow rate and/or a less
restrictive sample tube if the sampling method being used allows this.