Archimede Portable Dataloggers User Manual
rev.5 -03.12.2019-
The DL will suggest the first run to perform. At the end of each run, the DL will suggest the next
run. It is possible to interrupt the test at the end of a run by selecting
Save and exit
and pressing
ENTER
.
During the execution of a test, what is displayed in the central part of the screen depends on the
type of test. However, the upper part and the left part of the screen shows items common to all
tests.
In the upper part of the screen the
Name of the probe
being used and the
Name of the tube
are
shown. On the left the
Current date and time
and the
Measure unit
are shown. Four buttons are
also present:
Data
to view acquired data in tabular form or graphic form
Save
to save the acquired data and to exit
Read
to read the value currently displayed and proceed to the next step
Back
to return to the last step
It is possible to stop the test at any time. Pushing the
ESC
key, a confirmation request will appear.
If you confirm, the test will be deleted. Differently, holding the ESC key pushed for a few seconds,
the data will be saved and the test will be partial. You may also abort the test at any time by
pressing the
ESC
key: in this case no data will be saved.
During the inclinometer test, you could save a read using the external button (0CDL300REMC). This
button has the same function as the
Save
button. To prevent the user from inadvertently triggering
multiple readings, the external button is disabled for two seconds after each push. We advise to
wait at least three seconds between each pushing.
INCLINOMETER TESTS