Sirona Dental Systems GmbH
10 Operating light, tumbler filling, flushing of the cuspidor, purge
C3
+
10.3 Flushing of the cuspidor
59 06 404 D 3390
D 3390.201.01.18.02
61
10.3
Flushing of the cuspidor
When you press this key briefly, the flushing function is
switched ON and the cuspidor is flushed for a preset
period of time.
When you press the key again while the cuspidor is
being flushed, flushing stops.
Only for cuspidor flushing key of the dentist ele-
ment:
Pressing this key for > 3 seconds displays the settings
dialog
BOWL FLUSH TIMER
You can set a flushing time between 20 and
420 seconds (in steps) with the –
/
+ keys.
10.4
Purge / Auto Purge function (purging the water paths)
PURGE function
(only for dentist element)
The purge function allows for purging the water paths.
To do this, press the
quick adjustment key
1
(> 3 s)
when all instruments are in their holders.
You will be prompted to choose between
PURGE
/
APURGE
and
SANITATION
1
.
Select
PURGE
with the
–
or
+
key.
PURGE
must flash.
When you press
quick adjustment key 1
briefly (< 3 s),
the following prompt appears if all instruments are in
their holders:
SELECT ITEMS FOR PURGING
–
+
BOWL FLUSH TIMER
A
1
A
SELECT MODE
PURGE / APURGE / SAN
SELECT MODE
PURGE / APURGE / SAN
–
+
1. The SANITATION function is described in the document
"Care and Cleaning by the Practice Team".
A
SELECT ITEMS
FOR PURGING
SELECT ITEMS
FOR PURGING