Sirona Dental Systems GmbH
6 Foot control
C3
+
6.1 Wireless foot switch C+
59 06 404 D 3390
D 3390.201.01.18.02
33
1.
Remove the screws
(A)
from the bottom of the foot
switch.
2.
Remove the switch part
(C)
from the foot switch.
3.
Pull the battery out of the battery compartment by
pulling the cloth strap
(B)
and replace it with a new
one. Be careful to insert it with the correct polarity
(minus pole facing spring). The cloth strap
(B)
must
again lie underneath the battery.
Checking the battery status
The battery status is indicated by the yellow LED
(D)
in
the switch part.
•
LED lights up for approx. 10 s after battery is insert-
ed – Battery OK
•
LED flashes after battery is inserted – Battery too
weak, must be replaced
•
LED does not light up – Battery empty, must be re-
placed
Assembling the foot switch
1.
Mount the switch part
(C)
on the foot switch.
2.
Remove the screws
(A)
from the bottom of the foot
switch.
IMPORTANT
Restarting the wireless foot switch
Once you have changed the battery, switch the treat-
ment center back on and check that all functions of the
foot switch are working.
You do
not
need to register the foot switch at the treat-
ment center again after changing the battery.