Sirona Dental Systems GmbH
3 Technical description
C3
+
3.2 Technical data
59 06 404 D 3390
D 3390.201.01.18.02
19
3.2
Technical data
Dental treatment center
Model designation:
C3
+
Power connection:
100 V / 115 V / 230 VAC 50 / 60 Hz
Rated current:
11.5 A
at 100 VAC
9.5 A
at 115 VAC
4.5 A
at 230 VAC.
In addition, max. 6 A for external devices
Overvoltage category:
2 acc. to IEC 60664-1
Main unit fuse:
for 100/115 V AC: T 10 A H, 250 V AC , REF 10 77
460
for 230 V AC:
T 6.3 A H, 250 V AC , REF 10 77 452
Protection class:
Class I device
Device class in accordance
with Directive 93/42/EEC:
Class II
a
equipment
Degree of protection against elec-
trical shock:
Type
B
applied parts
Except SIROCAM 3 / SIROCAM C. These are:
Type
BF
applied parts
Degree of protection against
ingress of water:
Ordinary equipment (without protection against
ingress of water)
The foot switch has an IP X1 degree of protection
against liquids (drip-proof).
Mode of operation
Continuous operation with intermittent loading accord-
ing to the type of dental operation, see "Intermittent
operation" on page 14
This device is permanently installed. Operation is not
permitted in mobile vehicles.
Transport and storage conditions:
Temperature:
-40°C – +70°C
-40°F – +158°F
relative humidity:
10% – 95%
Air pressure:
500hPa – 1,060hPa
Operating conditions:
Ambient temperature:
10°C – 40°C
50°F – 104°F
relative humidity:
30% – 85% (no condensa-
tion)
Air pressure:
700hPa – 1,060hPa
Installation location:
≤
3,000 m above sea level
This treatment center is not suitable for operation in
areas subject to explosion hazards.
Pollution degree:
2 acc. to IEC 60664-1
Tests/Approvals:
See "Standards / Approvals" on page 17
Year of manufacture:
(on the rating plate of the patient chair)
-40
+70
1060
500
95
10
+10
+40
1060
700
85
30