H49.0.02.6C-02
Operating Manual MH 3750
Page 4
4/11
2 Device Configuration
For configuration of the device press "
Menu
"-key (key 4) for 2 seconds, the main menu will be shown (main display:
“SEt”). Choose the desired menu branch by pressing the "
Menu
"-key (key 4). By pressing “
4
” (key 3) the referring
parameters can be chosen. The referring values are changed by pressing the keys "
5
" (key 2) or "
6
" (key 5) (Choice
of parameter: “
4
”).
Pressing
“Menu”
(key 4) again will jump back to the main menu selection and stores the settings.
Use key "
Quit
" (key 6) to leave configuration.
Menu
Parameter Values
Meaning
key
Menu
key
4
key
5
or
6
p.r.t.
Set Configuration: Generic Settings
°C:
All temperature values are in degrees Celsius
°F:
All temperature values are in degrees Fahrenheit
*
0.1°:
Resolution 0.1°C
0.01°:
Resolution 0.01°C
Auto:
Resolution is selected automatically
* 4.1
E.751
Measuring / sensor curve according to EN60751
USEr
User sensor curve (Predefined to EN60751 values, changeable by
software MHKonfig)
*
4.2
-2.50°C..2.50°C resp.
-4.50°F...4.50°F
The zero point of the measurement will be displaced by this value to
compensate for deviations in the probe or in the measuring device.
oFF:
Zero displacement inactive (=0.0°)
*
4.3
-2.000...2.000:
The scale of the measuring will be changed by this factor (in %) to
compensate deviations of temperature probe or measuring device
oFF:
Scale correction factor inactive (=0.000)
*
4.4
1...120
Power-off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed/no interface communication takes place
oFF
Power-off function inactive (continuous operation, e.g. mains operation)
oFF:
Function of the output: No output function, lowest power consumption
SEr:
Output is serial interface
dAC:
Output is analogue output 0...1V
4.5
01,11..91
Base Address when Output = Serial Interface :
Base address of device for interface communication.
4.5.1
-200.0...850.0°C resp.
-328.0...1562.0°F
Output offset when output = analogue output: (depending on selected
unit) Input of the temperature at which 0V should be output
4.5.2
-200.0...850.0°C resp.
-328.0...1562.0°F
Output scale when output = analogue output: (depending on selected
unit) Input of the temperature at which 1V should be output
4.5.2
Set Alarm: Settings Of Alarm Function
on
Alarm
on, with horn-sound
no.So
Alarm on, without horn-sound
oFF
no alarm function
min range
...AL.Hi
Min alarm rail (not when AL. oFF)
AL.Lo
...max range
Max alarm rail (not when AL. oFF)
4.6
Set Logger: Configuration Of Logger Function
*
CYCL
Cyclic
: logger function ‚cyclic logger‘
Stor
Store
: logger function ‚individual value logger‘
oFF
no logger function
* 5
0:01..60:00
Cycle time of cyclic logger [minutes:seconds]
* 5.2
Set Clock: Setting Of Real Time Clock
HH:MM
Clock
: Setting of time hours:minutes
YYYY
Year
TT.MM
Date
:
day.month
4.7
Hint:
If keys ‘Set‘ and ‘Store‘ are pressed simultaneously for more than 2 seconds, the settings are reset to
the settings ex works
(*)
If the logger memory contains data already, the menus/parameters marked with (*) can not be invoked!
If these should be altered the logger memory has to be cleared before!