H49.0.02.6C-02

Operating Manual MH 3750

Page 11

11/11

8 Technical Data

Supported probes

Pt100 4-wire (2-wire possible)

Sensor Curve

according to EN60751
or with user-sensor curve (table of 50 rows and 2 columns: Ohm/°C]

Probe connection 

4pole Mini-DIN socket 

Resolution

0.01°C respectively 0.1°C, 0.01°F respectively 0.1°F 

Measuring Ranges 

0.01°C

0.1°C

0.01°F

0.1°F

–199.99... +199.99°C

-200.0... +850.0°C

-199.99... +199.99°F

-328.0... +1562.0°F

Precision 

Device without probe ±1Digit (at nominal temperature)

Range 0.01°C/F

Range 0.1°C/F

 ±0.03°C / 0.06°F  

±0.1°C / ±0.2°F

Measuring

4-wire measuring with automatic compensation of thermovoltage errors,
measuring current ca. 0.3mA

Temperature drift

<=0,002K per 1K for Pt100 4-wire 

Nominal temperature

25°C

Ambient

Temperature

-

25 ... +50°C (-13 .. 122°F)

rel. humidity 0 ... 95%r.F. (not condensing)

Storage temperature  -

25 ... +70°C (-13 ... 158°F)

Housing

Dimensions: 142 x 71 x 26 mm (L x W x D)
impact-resistant ABS plastic housing, membrane keyboard, transparent panel. Front side IP65,integrated 
pop-up clip for table top or suspended use.

Weight 

approx. 155 g

Output: 

3.5mm audio plug, stereo

 Selectable as serial interface: 

via optically isolated interface adapter GRS3100 or GRS3105 (p.r.t. accessories) directly 

connectable to RS232-interfaces.

   analogue output:

0..1V, freely scaleable (resolution 13bit, accuracy 0.05% at nominal temperature, cap. load <1nF)

Min/Max-Alarm:

The measuring value is constantly monitored for the min and max rails set. Alarming is done by integrated 
horn, display and interface

Real time clock:

Integrated clock with date and year

Logger:

2 Functions: individual value logger („Func–Stor“) and cyclic logger („Func–CYCL“)

 Memory: 

Stor: 99 data sets; CYCL: 16384 data sets 

 Cycle time CYCL:

0:01...60:00 (minutes:seconds, min 1s, max 1h)

Power Supply 

9V-Battery, type IEC 6F22 (included) as well as additional d.c. connector (diameter of internal pin 1.9 mm) 
for external 10.5-12V direct voltage supply.  (suitable power supply: GNG10/3000)

Power Consumption

output off

ca. 0.90mA

output serial interface: 

ca. 1.15mA

analogue output:

ca. 1.25mA

Display 

Two 4 ½  digits  LCD’s (12.4mm high and 7 mm high) for temperature,  min./ max values, hold function, etc. 
as well as additional pointing arrows.

Pushbuttons 

6 membrane keys for on/off switch, menu operation, min. and max. value memory, hold-function etc.

Min-/Max-Value Memory

 Both the max. and the min. value will be memorised.

Hold Function 

Press button to store current value.

Automatic-Off-Function 

Device will be automatically switched off if no key is pressed/no interface communication takes place for 

the time of the power-off delay. The power-off delay can be set to values between 1 and 120 min.; it can be 
completely deactivated.

EMC: 

The MH3750 corresponds to the essential protection ratings established in the Regulations of the Council 
for the Approximation of Legislation for the member countries regarding electromagnetic compatibility 
(89/336/EWG) 
EA1 +A2 (Appendix B, class B), additional error: < 1% FS. 

SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany

 

Tel: +49-5605-803-0, Fax: +49-5605-803-54, email: [email protected], web: www.SIKA.net

Содержание MH 3750

Страница 1: ...ert K hn GmbH Co KG Struthweg 7 9 34260 Kaufungen Germany Tel 49 5605 803 0 Fax 49 5605 803 54 email info SIKA net web www SIKA net Operating Manual Precision Thermometer MH 3750 from Version 1 0 For...

Страница 2: ...old to a warm environment condensation may result in a failure of the function In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start up 3 If...

Страница 3: ...nfiguration menu 1 3 Connections 1 Output Operation as interface Connect to optically isolated interface adapter accessory GRS 3100 or GRS3105 Operation as analogue output Connection via suitable cabl...

Страница 4: ...ed no interface communication takes place oFF Power off function inactive continuous operation e g mains operation oFF Function of the output No output function lowest power consumption SEr Output is...

Страница 5: ...heat loss caused by the probe is not considered When measuring fluids therefore the probe should be emerged sufficiently deep and be stirred continuously When measuring gases the probe should also eme...

Страница 6: ...ed Together with the zero displacement see above this factor is mainly used to compensate for sensor deviations Unless the factor is set to off the Corr arrow in the display shows an active scale corr...

Страница 7: ...ogue Output Scaling with DAC 0 and DAC 1 With the DAC 0 and DAC 1 values the output can be rapidly scaled to Your efforts Keep in mind not to connect low resistive loads to the output otherwise the ou...

Страница 8: ...w or by means of the interface and a PC with GSOFT3050 software Max number of measurings 99 A measuring contains current measuring value at the time of recording time and date of the recording After e...

Страница 9: ...lly will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by 5 key 2 and 6 key 5 Qui...

Страница 10: ...nnect sensor to socket sensor cable or device defective return to manufacturer for repair Value exceeding measuring range Check Is the value exceeding the measuring range specified temperature too hig...

Страница 11: ...is done by integrated horn display and interface Real time clock Integrated clock with date and year Logger 2 Functions individual value logger Func Stor and cyclic logger Func CYCL Memory Stor 99 da...

Отзывы: