Sigma 3-16PK Скачать руководство пользователя страница 17

CONTROLLED COPY 

 Centre for Biological Engineering 

 

Standard Operating Procedure  

 

 

 

 

 

 

 

SOP122 

 

Title:  Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK

 

 

Location: 

CBE Laboratory Unit (H21/22) 

 

Version 003   

Effective Date:  26

th

 June 2018 

Review 26

th

 June 2020 

 
Written by: Victoria Workman 
 

Reviewed by: P.Hourd 

Approved by: R/I.Temple 

 
 

Page 17 of 22 

 

Annex 1: Safe Use of Centrifuges (Extracts from BS4402) 

 

Moving parts in centrifuges can reach very high kinetic energies. Breakage of a rotor or accessories inside 
the centrifuge releases large amounts of energy, and debris can be expelled at high speeds from the 
centrifuge; bodily movement of centrifuges for a distance of several metres can also occur in this situation. 
 
Modern centrifuges must be designed and manufactured to British Standard BS 4402:1982, Safety 
Requirements for Laboratory Centrifuges. This Standard is intended to ensure that the casing of the 
centrifuge will contain the debris of components which break inside it and to ensure that centrifuges are 
fitted with necessary interlocks and other safety devices. 
 

It is important that all users of centrifuges are properly trained in the safe use and maintenance of 
the equipment. 
 

Appendix B of BS 4402 gives advice on the use of centrifuges and part of it reads: 
 
'RECOMMENDATIONS ON THE USE OF CENTRIFUGES’ 
 

B.1 Recommendations applying to all centrifuges 
 

B.1.1.  

Introduction

: When running, a centrifuge has considerable kinetic energy and even a small 'bench 

centrifuge' not complying with the requirement of this British Standard can expel debris, or even 
move bodily, for a distance of several metres in the event of mechanical failure. While considerable 
effort goes into the design and construction of centrifuges to try to minimise the hazards, sensible 
use is also important for the safety of operators and others in the laboratory. 

 
B.1.2  

Installation: 

It is essential that all centrifuges be fixed to the working surface, following the 

specifications and recommendations of the manufacturer for suitable fixings and fixing details. Care 
should be taken to ensure that the working surface and its substrate are strong enough to withstand 
the maximum forces of rotation and translation expected to act on the fixings in the event of a 
disruption of the rotation assembly (see instruction manual). A centrifuge may move abruptly in the 
event of such a disruption, in spite of the fixings, but the distance will be limited if the centrifuge is 
correctly fixed. 

 
B.1.3  

Training of Operators: 

Training in the operation of the centrifuge should be carried out by an 

experienced user or (by arrangement) by the manufacturer's representative, and should be given to 
all operators of the centrifuge before they are permitted to use the centrifuge. 

 
B.1.4  

Precautions to be taken when using centrifuges: 

NOTE: attention is also drawn to 

recommendations given in BS 5345 and to the 'User Guide for the Safe Operation of Centrifuges' 
published by the Institution of Chemical Engineers. 

 

B.1.4.1. Inspect each sample container and head accessory and all seals (if any) before filling. 

Reject any which are damaged or not authorised by the manufacturer, since such 
components could cause unsatisfactory sealing, breakage or jamming in the centrifuge head. 

 

Содержание 3-16PK

Страница 1: ...UST be used for applications involving the centrifugation of infectious toxic or pathogenic material SPECIAL NOTES HEALTH SAFETY 2 1 Hazards associated with use of centrifuge Mechanical failure of rot...

Страница 2: ...he start If a rotor other than the preselected one is installed the system will display the installed rotor and the pre selection speed will be corrected if necessary x Use only the correct designated...

Страница 3: ...ey intend to carry out iv Shall record alarms adverse events non conformances or malfunction on the Preventative Maintenance of Centrifuges Record Section 6 and notify the Laboratory Manager Responsib...

Страница 4: ...efore each use NOTE Lifetime of the Rotor After 10 years the rotors and accessories must be inspected by the manufacturers The rotor must be scrapped if it has done 50 000 cycles for reasons of safety...

Страница 5: ...up to the maximum recommended by the manufacturer according to the head and accessories to be used or to the recommended reduced speed if the sample has a high density viii Follow the manufacturer s i...

Страница 6: ...Always consider all components e g inserts and trunnion rings when balancing There should be no more than 100g difference between buckets next to each other iii Avoid contaminating the outside of a s...

Страница 7: ...2018 Review 26th June 2020 Written by Victoria Workman Reviewed by P Hourd Approved by R I Temple Page 7 of 22 iii Insert the holders into the rotor ensuring that the holders fit correctly into the t...

Страница 8: ...of the option Rotor on the display ii Select the option by pressing or turning the right rotary knob Set will now flash iii Turn the right rotary knob until the desired rotor number is displayed iv P...

Страница 9: ...isplayed iv Press the right rotary knob to confirm the selected value If this is not done the value will be automatically reset to the last setting v When the maximum permissible speed of the rotor is...

Страница 10: ...field in front of the parameter Temp ii Select the option by pressing or turning the right rotary knob Set will now flash iii Turn the right rotary knob until the desired value is displayed iv Press t...

Страница 11: ...l for continuous run v The program will only be loaded if the actual temperature is above the set temperature vi The program runs until the set value is reached Then a sound signal is issued and the p...

Страница 12: ...ted with seals in place vi If the temperature of the chamber is below room temperature pre cool the rotor to the lower temperature before securing the rotor this will minimize the chance of it seizing...

Страница 13: ...otate the rotor by hand without the rotor fastening nut If the motor shaft does not turn around on a perpendicular axis the motor and motor shaft must be replaced Refit the rotor correctly after visua...

Страница 14: ...l grain of rice to the load bearing bolts of the rotor and the buckets vi After completion of the routine clean record the event the Preventative Maintenance of centrifuges Record Section 6 Consult th...

Страница 15: ...lly removed Rubber inserts must be carefully cleaned or replaced If a problem has occurred the following must be considered Glass particles in the rubber cushion will cause glass breakage again Glass...

Страница 16: ...he CBE website NOTE A Declaration of decontamination available in the Operators Manual may also be required 5 7 Centrifuge or Rotor Decommissioning Checklist i If the centrifuge or rotor needs to be d...

Страница 17: ...British Standard can expel debris or even move bodily for a distance of several metres in the event of mechanical failure While considerable effort goes into the design and construction of centrifuge...

Страница 18: ...manufacturer s instructions for starting and where the control is not automatic for increasing the speed of rotation B 1 4 6 Allow the rotation assembly to come to rest before attempting to open the...

Страница 19: ...on of the rotor within the centrifuge The centrifugal forces cause rotors to stretch At low and normal speeds the change of size is elastic and the rotor returns to its original size when stationary A...

Страница 20: ...ach bucket Centrifuging Hazardous Materials Hazardous materials must be centrifuged in the proper type of sealed container The basic principle of safe containment is an air tight leak proof seal Most...

Страница 21: ...Unit H21 22 Version 003 Effective Date 26th June 2018 Review 26th June 2020 Written by Victoria Workman Reviewed by P Hourd Approved by R I Temple Page 21 of 22 Remember that it is important to follow...

Страница 22: ...rms from the SOP These are now on the CBE website iii Amendment to forms The QS Form 003 Preventative Maintenance of centrifuges now referenced in place of other inspection cleaning forms iv 5 5 3 Add...

Отзывы: