Sigma 3-16PK Скачать руководство пользователя страница 14

CONTROLLED COPY 

 Centre for Biological Engineering 

 

Standard Operating Procedure  

 

 

 

 

 

 

 

SOP122 

 

Title:  Use and Maintenance of Sigma Refrigerated Centrifuge 3-16PK

 

 

Location: 

CBE Laboratory Unit (H21/22) 

 

Version 003   

Effective Date:  26

th

 June 2018 

Review 26

th

 June 2020 

 
Written by: Victoria Workman 
 

Reviewed by: P.Hourd 

Approved by: R/I.Temple 

 
 

Page 14 of 22 

 

adaptors should be removed after use to allow adequate cleaning and drying. Ensure that the 
manufacturer's recommendations on cleaning materials are followed. 

 

 
EVERY MONTH: 

 

(i)  This is the responsibility of the responsible person ( or any laboratory user if the centrifuge is in 

need of cleaning) 
 

(ii)  Before cleaning, pull the mains plug  
 
(iii)  Wipe the centrifuge casing, venting slots, the rotor chamber, the rotor and the accessories with 

mild (pH 6 to 8) detergent (e.g. Fisherbrand or Neutracon).  

 
(iv)  During cleaning, make sure that cleaning liquids and especially organic solvents DO NOT come 

into contact with the drive shaft and the ball bearing 

(WARNING: organic solvents may

 

decompose the lubricant of the motor bearing and the drive shaft may block.)

  

 

(v)  After cleaning, dry the holders with tissue. Add a spot of grease the size of a small grain of rice to 

the load-bearing bolts of the rotor and the buckets 

 
(vi)  After completion of the routine clean, record the event the Preventative Maintenance of 

centrifuges Record. (Section 6). Consult the Responsible Person/Laboratory Manager if any 
signs of wear and tear, damage or corrosion are detected during the cleaning process. 

 

NOTE: 

This procedure increases life time and reduces corrosion. Corrosion or resultant damages 

which are caused by insufficient care do not constitute a warranty claim.  

 

5.5.5. Sterilization and Disinfection of Rotor Chamber and Accessories  
 

Where a spill or leak is detected within the centrifuge, the centrifuge MUST be disinfected immediately. 
The procedure will depend upon the risk group of the agent involved (refer to the Risk assessment). If 
there is reason to believe that a breakage may have occurred whilst the centrifuge was running, adopt 
the procedure in SOP038.  
 

 

NOTE: 

The centrifuges and the accessories consist of different materials. Before using detergents or 

decontamination agents other than those recommended by the manufacturer is applied, the user 
should contact the Laboratory Manager and/or the manufacturer to ensure that the planned procedure 
does not damage the equipment. 
 

If autoclaving:  

 

(i) 

Check the cleaning records for the centrifuge rotor BEFORE autoclaving. The rotor MUST NOT 
be autoclaved more than 10 times. 

Содержание 3-16PK

Страница 1: ...UST be used for applications involving the centrifugation of infectious toxic or pathogenic material SPECIAL NOTES HEALTH SAFETY 2 1 Hazards associated with use of centrifuge Mechanical failure of rot...

Страница 2: ...he start If a rotor other than the preselected one is installed the system will display the installed rotor and the pre selection speed will be corrected if necessary x Use only the correct designated...

Страница 3: ...ey intend to carry out iv Shall record alarms adverse events non conformances or malfunction on the Preventative Maintenance of Centrifuges Record Section 6 and notify the Laboratory Manager Responsib...

Страница 4: ...efore each use NOTE Lifetime of the Rotor After 10 years the rotors and accessories must be inspected by the manufacturers The rotor must be scrapped if it has done 50 000 cycles for reasons of safety...

Страница 5: ...up to the maximum recommended by the manufacturer according to the head and accessories to be used or to the recommended reduced speed if the sample has a high density viii Follow the manufacturer s i...

Страница 6: ...Always consider all components e g inserts and trunnion rings when balancing There should be no more than 100g difference between buckets next to each other iii Avoid contaminating the outside of a s...

Страница 7: ...2018 Review 26th June 2020 Written by Victoria Workman Reviewed by P Hourd Approved by R I Temple Page 7 of 22 iii Insert the holders into the rotor ensuring that the holders fit correctly into the t...

Страница 8: ...of the option Rotor on the display ii Select the option by pressing or turning the right rotary knob Set will now flash iii Turn the right rotary knob until the desired rotor number is displayed iv P...

Страница 9: ...isplayed iv Press the right rotary knob to confirm the selected value If this is not done the value will be automatically reset to the last setting v When the maximum permissible speed of the rotor is...

Страница 10: ...field in front of the parameter Temp ii Select the option by pressing or turning the right rotary knob Set will now flash iii Turn the right rotary knob until the desired value is displayed iv Press t...

Страница 11: ...l for continuous run v The program will only be loaded if the actual temperature is above the set temperature vi The program runs until the set value is reached Then a sound signal is issued and the p...

Страница 12: ...ted with seals in place vi If the temperature of the chamber is below room temperature pre cool the rotor to the lower temperature before securing the rotor this will minimize the chance of it seizing...

Страница 13: ...otate the rotor by hand without the rotor fastening nut If the motor shaft does not turn around on a perpendicular axis the motor and motor shaft must be replaced Refit the rotor correctly after visua...

Страница 14: ...l grain of rice to the load bearing bolts of the rotor and the buckets vi After completion of the routine clean record the event the Preventative Maintenance of centrifuges Record Section 6 Consult th...

Страница 15: ...lly removed Rubber inserts must be carefully cleaned or replaced If a problem has occurred the following must be considered Glass particles in the rubber cushion will cause glass breakage again Glass...

Страница 16: ...he CBE website NOTE A Declaration of decontamination available in the Operators Manual may also be required 5 7 Centrifuge or Rotor Decommissioning Checklist i If the centrifuge or rotor needs to be d...

Страница 17: ...British Standard can expel debris or even move bodily for a distance of several metres in the event of mechanical failure While considerable effort goes into the design and construction of centrifuge...

Страница 18: ...manufacturer s instructions for starting and where the control is not automatic for increasing the speed of rotation B 1 4 6 Allow the rotation assembly to come to rest before attempting to open the...

Страница 19: ...on of the rotor within the centrifuge The centrifugal forces cause rotors to stretch At low and normal speeds the change of size is elastic and the rotor returns to its original size when stationary A...

Страница 20: ...ach bucket Centrifuging Hazardous Materials Hazardous materials must be centrifuged in the proper type of sealed container The basic principle of safe containment is an air tight leak proof seal Most...

Страница 21: ...Unit H21 22 Version 003 Effective Date 26th June 2018 Review 26th June 2020 Written by Victoria Workman Reviewed by P Hourd Approved by R I Temple Page 21 of 22 Remember that it is important to follow...

Страница 22: ...rms from the SOP These are now on the CBE website iii Amendment to forms The QS Form 003 Preventative Maintenance of centrifuges now referenced in place of other inspection cleaning forms iv 5 5 3 Add...

Отзывы: