Version 06/2007, Rev. 1.15 of 16/09/2011 - sb
page 69 of 75
Translation of the original operating manual
11.2 Table of rotors and accessories with a different service life
Rotor /
bucket
Cycles
Service life
("Exp.Date")
Autoclaving
Suitable for centrifuge
Remarks
11026
7 years
1-14, 1-14K
12082
7 years
1-14, 1-14K
12083
7 years
1-14, 1-14K
12084
7 years
1-14, 1-14K
12085
7 years
1-14, 1-14K
12092
5 years
20x
1-14, 1-14K
12093
5 years
20x
1-14, 1-14K
12094
5 years
20x
1-14, 1-14K
12096
5 years
20x
1-14, 1-14K
12101
5 years
20x
1-15, 1-15K,
1-15P, 1-15PK
12124
5 years
20x
1-15, 1-15K,
1-15P, 1-15PK
12126
5 years
20x
1-15, 1-15K,
1-15P, 1-15PK
9100
15,000
4-15C, 4K15C, 4-16, 4-16K,
6-15, 6K15, 6-16, 6-16K
without engraving,
only "spincontrol
professional"
12500
7 years
6-15, 6K15, 6-16, 6-16K
13218
20,000
4-16, 4-16K, 6-16, 6-16K
13845
20,000
8K
13850
10,000
8K
13860
35,000
8K
13864
1,000
8K
without engraving
13865
1,000
8K
without engraving
Rotors and accessories with a
different service life
If other no data concerning the service life are engraved on the rotor or accessory, rotors
and buckets must be checked by the manufacturer after 10,000 cycles. After 50,000 cycles,
rotors must be scrapped for safety reasons.