Laboratory Centrifuge Sigma 2-16P
6 Using the centrifuge
38 / 76
Version 04/2007, Rev. 2.23 of 28/04/2022
• sb
Translation from the original operating manual, part no. 0700602
Continuous run
During the continuous run, the runtime of the centrifuge is unlimited and
must be stopped manually. The centrifuge accelerates during the
continuous run until the set speed is reached.
•
Select the modification mode and turn the function knob from the time
00:10 anti-clockwise or from the time 11:59 clockwise. The indication
"Cont" and the symbol "
∞" will be displayed in the time field.
Fig. 12:
Indication “Cont” during a continuous run
Short run
• Keep the start key pressed during the short run.
During the short run, the centrifuge accelerates at maximum power until the
maximum speed is reached. The runtime is counted and in the time field
the indication "short run" is displayed.
When the start key is released, the centrifuge decelerates at maximum
power to a standstill.
Fig. 13:
Indication “short run” during a short run
Changing the time increments
By default, the runtime is changed in intervals of 10 sec (in the mode
min:sec) or in intervals of 10 min (in the mode hrs:min)
To change the runtime in intervals of 1 sec or 1 min:
• Keep the stop key pressed while setting the desired runtime with the
right function knob
Pos : 129 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1