Laboratory Centrifuge Sigma 2-16P
6 Using the centrifuge
32 / 76
Version 04/2007, Rev. 2.23 of 28/04/2022
• sb
Translation from the original operating manual, part no. 0700602
Pos : 106 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0021 Eins et zen eines Wi nk elrot ors mit H ermetikdeck el 3 Nm @ 29 \mod_1405321531516_68. docx @ 197015 @ 4 @ 1
→
6.2.2.2
Installation of an angle rotor with a hermetic lid
• Open the centrifuge lid by pressing the lid key.
• Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
• Screw the rotor lid onto the rotor and tighten it by turning it clockwise.
• Lower the rotor with the lid with its central bore straight down onto the
motor shaft.
• Insert the rotor tie-down screw into the motor shaft. Tighten the rotor tie-
down screw clockwise with the supplied rotor wrench with 3 Nm. In
doing so, hold the rotor at its outer rim.
• Follow the safety instructions and hazard warnings (see chapter 3 -
NOTE
The rotor can also be used without a cover.
NOTE
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft.
Removal of a rotor
• Unscrew the rotor tie-down screw (counter-clockwise) and remove the
rotor with the lid.
• Open the lid by turning it counter-clockwise and remove the lid.
1
Rotor tie down screw
2
Lid
3
Rotor
Fig. 5: Angle rotor with a hermetic lid
Pos : 107 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1