Laboratory Centrifuge Sigma 2-16P
6 Using the centrifuge
Version 04/2007, Rev. 2.23 of 28/04/2022
• sb
37 / 76
Translation from the original operating manual, part no. 0700602
Pos : 119 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020 M anueller Betri eb @ 28\m od_1405320526938_68. docx @ 194439 @ 3 @ 1
→
6.3.2
Manual mode
Pos : 120 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020-0010 St art en ei ner Z entrif ugati on @ 28\m od_1405320527735_68. docx @ 194453 @ 4 @ 1
→
6.3.2.1
Starting a centrifugation run
The centrifuge is ready for operation when the start key is illuminated.
• Press the start key in order to start a centrifugation run.
Pos : 121 /010 Univ ers almodul e/ Leerzeil e 8pt @ 12\m od_1337063062114_0.docx @ 65687 @ @ 1
Pos : 122 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020-0020 U nt erbr ec hen einer Z entrifug ation @ 28\mod_1405320528561_68. doc x @ 194467 @ 4 @ 1
→
6.3.2.2
Interrupting a centrifugation run
• Press the stop key in order to interrupt a centrifugation run.
The centrifugation run will be terminated prematurely.
Pos : 123 /010 Univ ers almodul e/ Leerzeil e 8pt @ 12\m od_1337063062114_0.docx @ 65687 @ @ 1
Pos : 124 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020-0030 U nt erbr ec hen eines Brems vorgangs @ 28\m od_1405320529412_68.docx @ 194481 @ 4 @ 1
→
6.3.2.3
Interrupting a deceleration process
• Press the start key during a deceleration process in order to interrupt it
and to restart the centrifuge.
Pos : 125 /010 Univ ers almodul e/ Leerzeil e 8pt @ 12\m od_1337063062114_0.docx @ 65687 @ @ 1
Pos : 126 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020-0040 Auswahl, Anzeig e und Änderung von D at en @ 28\m od_1405320530250_68.docx @ 194495 @ 4 @ 1
→
6.3.2.4
Selection, display, and modification of data
• Turn the left function knob in order to select a field. The selected field
will be preceded by the indication "set", "run", or "save".
• Press the left function knob. The indication ("set", "run", or "save")
flashes and the modification mode is active.
• Turn the right function knob in order to modify the set value of the
selected field.
• Press the fright unction knob again to confirm the input and to quit the
modification mode
• If the input is not confirmed, the value will be automatically reset to the
last setting.
Pos : 127 /010 Univ ers almodul e/ Leerzeil e 8pt @ 12\m od_1337063062114_0.docx @ 65687 @ @ 1
Pos : 128 /100 Sigma/100 BA Z entrif ug en Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020-0050 Laufzei t - ohne T emp @ 28\mod_1405320531525_68. docx @ 194509 @ 4 @ 1
→
6.3.2.5
Runtime
The runtime is displayed in the left section of the display and can be set at
different intervals in a range from 10 seconds to 11 hours and 59 minutes.
The parameter runtime can be changed during the centrifugation.
NOTE
If the runtime is changed during an active centrifugation run, the time that
has already elapsed will not be taken into consideration. The centrifuge
will perform a complete run with the new runtime.
Fig. 11: Setting the runtime