Refrigerated Centrifuge with Heating System Sigma 4-16KHS
6 Using the centrifuge
Version 10/2012, Rev. 1.22 of 08/05/2020
• sb
49 / 91
Translation of the original operating manual, part no. 0701802
Start delay
If the start delay function is active, the centrifuge will not start until the
preset time has elapsed. The symbol "
" will be displayed in the runtime
field.
Display deceleration time
In the activated mode, the deceleration time will be displayed instead of the
runtime during and after the deceleration process. Below the time display,
the symbol "
" will be displayed. During a deceleration process, the
symbol flashes. Once the deceleration is complete, it is displayed in a
permanent manner.
Display
Temperature unit
The temperature can be displayed in °C (Celsius) or °F (Fahrenheit).
Language
The control system can be used in various language versions.
If a language is selected by mistake, it can be changed on any screen as
follows:
• Press and hold the stop button.
• Turn the function knob one notch to the left and then one notch to the
right.
• Release the stop button. The "Language" window will be displayed.
• Select the desired language.
Speed/RCF fine and Time fine
This menu item can be used to preselect the set speed in steps of 1 rpm
(instead of 100 rpm), the RCF value in steps of 1 x g (instead of 10 x g),
and the set time in steps of 1 min or 1 sec (instead of 10 min or 10 sec).
NOTE
Regardless of the fine adjustment, the step size increases when the
function knob is turned quickly.
Invert
If this function is activated, the display switches from the standard setting
with a bright background and dark writing to a dark background with bright
writing.