Refrigerated Centrifuge with Heating System Sigma 4-16KHS
3 Safety
Version 10/2012, Rev. 1.22 of 08/05/2020
• sb
21 / 91
Translation of the original operating manual, part no. 0701802
Pos : 55 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0060-0070 Sicherheit von R ot oren und Zubehör @ 44\m od_1470119200568_68.docx @ 437008 @ 3 @ 1
3.6.7
Safety of rotors and accessories
Pos : 56 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0060-0070- 0010 Lebens dauer (außer 8KBS) @ 26\m od_1405319032612_68. docx @ 192143 @ 3 @ 1
3.6.7.1
Service life
The rotors and accessories have a limited service life.
WARNING
Perform regular checks (at least once per month) for safety reasons!
• Pay special attention to changes, such as corrosion, cracks, material
abrasion, etc.
• After 10 years, they must be inspected by the manufacturer.
• After 50,000 cycles, the rotor must be scrapped for reasons of safety.
• If other data concerning the service life are engraved on the rotor or
bucket, these data shall apply accordingly. For example, a bucket with
the engraving "max. cycles = 10,000" has a service life of 10,000
cycles, and a rotor with the engraving "Exp. date 01/27" must be
scrapped in January 2027 at the latest (see figure).
• If a specification concerning the maximum number of cycles
and
a
specification concerning the service life (i.e. a date) are provided, the
specification that occurs first shall apply.
Fig. 4: Different service life
– engraving on the bucket/rotor
NOTE
• Refer to the table of rotors and accessories with a different service
life (see chapter 11.4 - " Table of the service life of rotors and
accessories ")!
Pos : 57 / 010 U nivers almodule/ Seitenw ec hsel @ 0\m od_1202116244312_0. doc x @ 105 @ @ 1