Refrigerated Centrifuge with Heating System Sigma 4-16KHS
5 Set-up and connection
28 / 91
Version 10/2012, Rev. 1.22 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0701802
Pos : 99 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 050 Aufst ellung und Ans chl uss/ 050- 0020-0020 Sic herungen baus eits ( außer 3-30KS) @ 7\mod_1306312302397_68. docx @ 43513 @ 3 @ 1
→
5.2.2
Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp L or B fuses that
are to be provided by the customer.
NOTE
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.