![Sigma 147161 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/sigma/147161/147161_operating-manual_1267884034.webp)
Refrigerated Centrifuge with Heating System Sigma 4-16KHS
6 Using the centrifuge
34 / 91
Version 10/2012, Rev. 1.22 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0701802
Pos : 117 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0040 Adapt er - nur L+S @ 27\mod_1405319210191_68. docx @ 192563 @ 4 @ 1
→
6.2.2.4
Adapters
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
• Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
• If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.
Pos : 118 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 119 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0051 Gefäß e ( 4-16S, 4-16KS, 4-16KHS) @ 56\mod_1521541840863_68. doc x @ 559986 @ 4 @ 1
→
6.2.2.5
Vessels
• Load the vessels outside of the centrifuge. Liquids in the buckets or
multiple carriers cause corrosion.
• Fill the vessels carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
• Always fill the tubes up to their useful volume (= the volume that is
stated for the tube).
• Follow the safety instructions and hazard warnings (see chapter 3 -
Maximum speed for tubes
Some tubes, such as centrifuge glass tubes, microtubes, culture tubes,
fluoropolymer tubes and especially high-volume tubes can be used in our
rotors, buckets, and adapters at higher speeds than their breaking limit.
NOTE
When using glass vessels, the maximum value of 4,000 x g must not be
exceeded (except special high-strength glass tubes; please refer to the
information provided by the manufacturer).
CAUTION
At speeds above 8,000 rpm, there is an increased risk of breakage, in
particular for 250 ml bottles.
Pos : 120 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1