
3EDU
● If the device is set for pressure measurement, no other measured variables (e.g. volume
flow) are available.
Example
For the pressure measurement, you set the following values, for example:
Damping value:
2.0 s
Lower range value:
0.0 bar
Upper range value:
5.0 bar
Application:
Pressure: linear (PRESS)
Unit:
bar
Lower scaling point:
-
Upper scaling point:
-
9.2.4.2
Level measurement
To set the application of the device for the level measurement, select the "level" characteristic
curve using the "Application" parameter (LEVEL).
For a level measurement, the device calculates the level height and the hydrostatic pressure.
The geometry of the vessel is not included in the calculation.
● The device uses a linear characteristic curve:
Parameter assignment
9.2 Parameter assignment over device with display
SITRANS P320/P420 (mA/HART)
Operating Instructions, 06/2018, A5E44852162-AA
109
Содержание SITRANS P320
Страница 10: ...Table of contents SITRANS P320 P420 mA HART 10 Operating Instructions 06 2018 A5E44852162 AA ...
Страница 18: ...Introduction 2 7 Notes on warranty SITRANS P320 P420 mA HART 18 Operating Instructions 06 2018 A5E44852162 AA ...
Страница 174: ...Functional Safety 10 7 Repair and service SITRANS P320 P420 mA HART 174 Operating Instructions 06 2018 A5E44852162 AA ...
Страница 260: ...Index SITRANS P320 P420 mA HART 260 Operating Instructions 06 2018 A5E44852162 AA ...