SITRANS P DS III/P410 (7MF4.33.. 7MF4.34.. 7MF4.35..)
A5E03434626-04, 05/2015
43
Differential pressure and flow rate input
16 … 1600 mbar
1.6 ... 160 kPa
6.424 … 642.4 inH
2
O
50 … 5000 mbar
5 ... 500 kPa
20.08 … 2008 inH
2
O
0.3 … 30 bar
0.03 ... 3 MPa
4.35 … 435 psi
6 … 600 mbar
0.6 ... 60 kPa
2.409 … 240.9 inH
2
O
420 bar
42 MPa
6091 psi
16 … 1600 mbar
1.6 ... 160 kPa
6.424 … 642.4 inH
2
O
50 … 5000 mbar
5 ... 500 kPa
20.08 … 2008 inH
2
O
0.3 … 30 bar
0.03 ... 3 MPa
4.35 … 435 psi
Lower measuring limit
•
Measuring cell with silicone oil filling
-100 % of max. measuring range
(-33 % for 30 bar/3 MPa/435 psi measuring cell) or 30 mbar a/3 kPa
a/0.44 psi a
Upper measuring limit
100% of maximum span
Start of scale value
Between the measuring limits (fully adjustable)
1)